8. If the full scale calibration is below 103, the display will be
unable to calculate a valid calibration, and will show cErr
(calibration error) after a few seconds. The existing calibration
will be retained, and the display will return to normal operation.
To view the existing calibration:
1. Press either the NEXT MENU or UP/ENTER button (but not
both) on the back panel, the display will show the existing
full scale analog calibration while the button is held down.
Release the button to return to normal operation.
To test the analog output:
1. While either the NEXT MENU or UP/ENTER button on the
back panel is pressed, the display will show the full scale
calibration and the analog output will go to full scale (20 mA).
This can be used to test or calibrate the equipment connected
to the analog output.
2. While the display is in the calibration mode (entered by
pressing both the NEXT MENU and UP/ENTER buttons) the
analog output will be at 4mA.