Download Print this page

Bosch EPS 200 Operating Instructions Manual page 9

Tester for holder assemblies and injectors
Hide thumbs Also See for EPS 200:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 8
7.2.�
Replacing the test oil
!
Before replacing the test oil, the EPS 200 must be
turned off and disconnected from the power supply.
Authorized types of test oil:
Oil type
SHELL Calibrati-
on Fluid S 9365
SHELL V-OIL 1404
SHELL Normafluid B. R.
VISCOR
Calibration Fluid 1487 AW-2
CASTROL
fluido para Calibracao 4113
ESSO EGL 70 147
BENZ UCF-1 Calibration Fluid
Procedure:
1. Turn off the EPS 200.
2. Disconnect the EPS 200 from the power supply.
3. Unscrew the fastening screws (x8) on the left of the
housing cover (Fig. 1; 6).
4. Remove housing cover (Fig. 1; 5).
5. Unscrew the ground lead on the side panel of the
housing (Fig. 1; 7).
!
As soon as the quick release coupling is attached
to the EPS 200, test oil flows through the hose
line. Always place the end of the hose line in the
collecting tank first.
6. Place the end of the 1.5 m hose line (Fig. 1, 4;
1 680 712 283) without the quick release coupling in
the collecting tank (Fig. 1, 3).
Robert Bosch GmbH
Manufacturer
Shell International
Shell Germany
SHELL France
Rock Vallery
Castrol Brazil
Esso AG
Benz Oil
Maintenance | EPS 200 | �
7. Insert the 1.5 m hose line (1 680 712 283) with quick
release coupling into the quick release couplings on
the test oil tank (Fig. 1; 1).
? Test oil flows into the collecting tank.
7
8
6
5
4
Fig. 1:
Draining the test oil
8. Once the test oil has drained out of the test oil tank,
insert the hose line into the quick release coupling
on the settling chamber (Fig. 1; 2).
? Test oil flows from the settling chamber into the
collecting tank.
9. Once the test oil has drained out of the settling
chamber, insert the hose line into the quick release
coupling on the oil cooler (Fig. 1, 8).
? Test oil flows from the oil cooler into the
collecting tank.
10. Once the test oil has drained out of the oil cooler,
disconnect the hose line from the quick release
coupling.
11. Screw the ground lead onto the housing cover.
12. Replace the housing cover and screw into place.
!
If the test oil is added too quickly, it is possible that
it will run out of the upward facing flushing outlet or
the connection for the external extractor (Refer to
Fig. 1, Item 5 in the EPS 200 operating instructions).
13. Pour the new test oil (ISO 4113) into the jet cham-
ber (Refer to Fig. 3, Item 2 in the EPS 200 operat-
ing instructions) very slowly and carefully. The test
oil tank holds around 6.5 liters.
14. Add test oil until it reaches the upper marking in
the oil level display (Refer to Fig. 1, Item 9 in the
EPS 200 operating instructions).
"
Replacement of the test oil is complete.
EPS 200 | �
| �
en
2
1
3
|
1 689 989 050
2009-03-17

Advertisement