Jäger ADwin-Gold-USB Manual page 25

Hide thumbs Also See for ADwin-Gold-USB:
Table of Contents

Advertisement

ADwin
The program outputs the values 0...65535 digits on both DACs, compares them
to the measured input values and displays the deviation in graphs.
The deviation should be less than 5 digits.
With Close you return to the overview window.
With "Print Calibration" you can print a protocol of the specified calibra-
tion data.
In the open window you can enter different information which will be presented
in your printout (for a later allocation to the protocol). With "Print" you start
printing; the program automatically returns to the overview window.
In the protocol you will find the calibration settings of all inputs and outputs for
gain and offset as well as the date of print.
The calibration is finished.
ADwin-Gold USB / ENET, manual version 3.8, October 2005
Calibration
Step 4
Step 5
21