Inputs and Outputs
analog
inputs/outputs
ANALOG OUT
ANALOG IN
CO POWER IN
DIO 00-15 (IN)
DIO 16-31 (OUT)
digital
inputs/outputs
10
CO1
ADwin-
POWER
GND
GND/PE
connector
Fig. 5 – Schematic of ADwin-Gold-D (ENET version)

5.1 Analog Inputs and Outputs

In order to operate the system without any interferences, isolated BNC con-
nectors are necessary. Otherwise there will be the danger of damages caused
by ESD or short circuits at the inputs. This will be the case when using not iso-
lated BNC T-pieces.
The ADwin-Gold device has to be connected to earth, in order to execute
measurement tasks without any interferences. Connect the GND plug via a
low-impedance solid-type cable with the central earth connection point of your
device.
The power supply from the power adapter at the computer also connects the
earth of the ADwin-Gold system with the earth of the computer. If you do not
operate the PC and the ADwin-Gold system in the same place, you should not
use the power supplied by the PC but an external power supply unit which is
earth-free, in order to avoid influences by different ground reference potentials.
In addition to the description of the inputs and outputs you will find notes below
for the conversion of digits into voltage values and for the input settings of the
analog inputs.
The pin assignment of the analog channels for ADwin-Gold-D is shown in
Fig. 6.
additional
inputs/outputs
CO2
CO3
Gold
PC
power-
ENET
supply
connector
ADwin-Gold USB / ENET, manual version 3.8, October 2005
ADwin
CO4
CAN 1.1
CAN 1.2
CAN 2
COM1
COM2

Advertisement

Table of Contents
loading

This manual is also suitable for:

Adwin-gold-enet

Table of Contents