Alarm Conditions - Delfi PTS ii Operator And Maintenance Manual

Portable tourniquet system
Hide thumbs Also See for PTS ii:
Table of Contents

Advertisement

2.17 
ALARM CONDITIONS 
There  are  a  number  of  conditions  for  which  the  PTS  ii 
will  produce  visual  and  audible  alarms.  Those 
conditions,  indications  and  appropriate  actions  are 
shown in Table 2.1.  The appropriate actions indicated 
are based on the most probable causes and should only 
be  used  as  a  guide.    Other  causes  of  alarm  conditions 
may indicate a need for other actions.   
The  alarm  indicator  light  is  illuminated and an audible 
alarm tone is produced whenever a alarm condition is 
detected.  
2.20.1  ALARM SILENCE FUNCTION 
Most  audible  alarm  tones  may  be  silenced  for  30 
seconds  by  momentarily  depressing  the  alarm  silence 
button which is displayed whenever a silencable alarm 
condition is present. At the end of the silenced period, 
the  alarm  tone  will  restart  if  the  alarm  condition  has 
not  been  corrected.    The  alarm  tone  may  be  silenced 
for additional 30 second periods as required. 
2.20.2  PRESSURE HIGH or LOW ALARMS 
A  "HIGH  PRESSURE"  or  "LOW  PRESSURE"  alarm  will 
occur  when  the  pressure  in  a  cuff  is  more  than  15 
mmHg  from  the  pressure  set  point.    To  minimize 
nuisance  alarms  that  can  be  caused  by  vigorous 
movement of the patient's limbs, a 1 second delay has 
been  designed  into  alarm  actuation  for  these 
conditions. 
2.20.3.  LEAK ALARMS 
It  is  possible  for  the  system  to  have  a  substantial  leak 
that  the  unit  can  compensate  for  by  continual 
pumping.    This  type  of  leak  could  be  due  to  a  hole  in 
the  cuff  or  hose  assembly,  a  loose  or  worn  hose 
connector,  or  leaks in the pneumatic circuit inside the 
PTS  ii  unit.    All  leaks  require  immediate  attention, 
because  they  could  progress  into  a  total  failure  of  a 
cuff to hold pressure at any time.  To alert the operator 
that  a  substantial  leak  is  present,  the  "CUFF  LEAK" 
alarm is activated when this type of leak is continuously 
present  for  more  than  7  seconds,  even  if  the  unit  is 
maintaining  the  cuff  pressure  within  15  mmHg  of  the 
set point.  If a "CUFF LEAK" alarm occurs, the cuff, hose 
assembly,  and  hose  connectors  should  be  checked  for 
leaks.  If an external leak cannot be found, test the PTS 
ii unit per Section 3.4 below.2.8.2. 
2.20.4  INTERNAL HARDWARE FAILURES 
When  "SYSTEM  ERROR"  and  a  numeric  error  code 
appear in the pressure and time displays during use or 
power‐up,  an  internal  hardware  failure  has  likely 
occurred  and  the  PTS  ii  unit  is  unusable.    In  this 
situation,  it  is  likely  that  the  unit  has  put  itself  in  the 
'safe  state'  mode,  in  which  the  pneumatic  valve  and 
pump are disabled and the current pressure in the cuff 
is held (in the absence of leaks).  It is also likely that a 
tone  will  sound  under  these  conditions.    The  'safe 
state' mode helps prevent unexpected loss of occlusion 
during a procedure if a sudden failure occurs. 
Although  it  is  very  unlikely,  internal  hardware  failures 
may  also  cause  erratic  operation  and/or  unintelligible 
displays  with  or  without  alarms,  and  may  or  may  not 
put the PTS ii in the 'safe state' mode.   
If either type of error occurs: 
a)  Set the unit to standby by pressing the 
'ON/STANDBY' button.  This removes power from 
the internal instrument circuitry and all instrument 
functions, causing the cuff to hold pressure (in the 
absence of leaks). 
b)  If required, attempt to restart the unit by pressing 
the 'ON/STANDBY' again to restart the unit.   
c)  If  required  to  continue  the  procedure,  clamp  the 
hose  with  a  hemostat  to  maintain  cuff  pressure, 
then immediately disconnect the faulty PTS ii unit 
and connect a replacement unit. 
d)  If  cuff  deflation  is  required,  disconnect  the  cuff 
from the PTS ii unit. 
WARNING
: In all cases of internal hardware failure, 
erratic  operation,  or  unintelligible  displays,  it  is 
possible that the pressure in the cuff is not accurately 
displayed  on  the  PTS  ii  unit  and  that  cuff  pressure 
may be present even when the PTS ii unit appears to 
be shut down.  The user must immediately determine 
if  the  cuff  is  inflated  or  deflated  and  continually 
monitor the cuff to ensure patient safety.  If deflation 
of the cuff is necessary, disconnect the hose from the 
PTS  ii  unit or from the cuff and confirm that the cuff 
deflates completely. 
18

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents