Delfi PTS ii Operator And Maintenance Manual page 23

Portable tourniquet system
Hide thumbs Also See for PTS ii:
Table of Contents

Advertisement

 
 
CHECK PRINTER 
The unit has detected that the 
printer is not ready to print.   
 
 
 
NO PRINTER 
The unit can not communicate 
with the optional printer. 
 
 
 
BATTERY FAIL 
The battery voltage has dropped 
below the threshold of safe 
operation and has shut down in 
a safe state* 
OR 
The battery has been 
removed/disconnected. 
Alarm message alternates with elapsed 
time display
Alarm message alternates with elapsed 
time display
22
 
Check that the printer is not out of labels; 
the labels are correctly installed; and the 
printer door is fully closed.
 
 
 
The unit was unable to communicate 
with the optional printer.  Check to make 
sure the printer is correctly connected to 
the PTS ii, and powered on. 
 
 
 
 
 
Press 'ON/STANDBY' to shut unit down.  
When cuff deflation is required, 
disconnect hose to cuff.  Ensure that the 
battery is connected properly.  Plug in AC 
power to attempt to recharge the 
battery.  Service or replace the battery 
pack (see Section 3.5 below).
 
 
 