2.8

ALARM CONDITIONS

There are a number of conditions for which the P.T.S. will produce visual and audible alarms. Those conditions, indications and appropriate actions are
shown in Table 2.1. The appropriate actions indicated are based on the most probable causes and should only be used as a guide. Other causes of
alarm conditions may indicate a need for other actions.
Condition
Description
High
The unit has detected
Pressure
high pressure in the
above set
cuff. High pressure is
point
defined as pressure
that is +15mmHg
above the pressure set
point continuously for
more than 1 second.
Low
The unit has detected
Pressure
low pressure in the cuff.
below set
Low pressure is defined
point
as pressure that is
15mmHg below the
pressure set point
continuously for more
than 1 second.
Table 2.1: Alarm Conditions
Message
Alarm
Pressure
display
tone
display
(steady)
PRESSURE
Constant
Current
HIGH
cuff
pressure,
flashing
PRESSURE
Constant
Current
LOW
cuff
pressure,
flashing
Action to silence
Time
alarm tone for 30
display
seconds
Normal
Momentarily Press
'PRESSURE + ',
'PRESSURE – ',
'TIME + ', or
'TIME – '.
Normal
Momentarily Press
'PRESSURE + ',
'PRESSURE – ',
'TIME + ', or
'TIME – '.
Action required to correct
alarm condition
Press and hold 'PRESSURE + '
to adjust set pressure to within
15mmHg of sensed pressure
OR
Press and hold 'DEFLATE' for 2
seconds to deflate the cuff.
Alarm will stop automatically
whenever the unit can regulate
the cuff to within 15 mmHg of the
pressure set point.
Press and hold 'PRESSURE – '
to adjust set pressure to within
15mmHg of sensed pressure
OR
Press and hold 'DEFLATE' for 2
seconds to deflate the cuff.
Alarm will stop automatically
whenever the unit can regulate
the cuff to within 15 mmHg of the
pressure set point.