2.5
PRESSURE SET POINT ADJUSTMENT
a) At the main screen (Figure 2.5), momentarily press
the pressure adjustment button. The PTS ii will
display the pressure adjustment screen (Figure
2.6). A pressure gauge icon is displayed and the
pressure set point is highlighted to indicate that
the set point is being adjusted.
b) Use the increment and decrement buttons to
modify the pressure set‐point value.
Momentarily pressing the increment button will
increase the set point by 5 mmHg. When the
button is held the set point will increase by 10
mmHg every second. Momentarily pressing the
decrement button will decrease the set point by 5
mmHg. When the button is held the set point will
decrease by 10 mmHg every second.
c) Press the back button to return to the main screen.
The PTS ii will automatically return to the main
screen after 3 seconds if no buttons are pressed.
Figure 2.6
Pressure Set Point Adjustment
d) Note to increase the pressure set point above
475mmHg confirmation is required (Figure 2.7).
Momentarily press the OK button to enter
extended pressure mode or the press back button
to return without entering extended pressure
mode. If confirmed pressures above 475 mmHg
may be selected. The pressure set‐point value and
cuff pressure are shown with a yellow background
when a pressure set point greater than 475 mmHg
has been selected.
Extended Pressure Mode
2.6
TIME LIMIT SET POINT ADJUSTMENT
a) At the main screen (Figure 2.5) momentarily press
the time limit adjustment button. The PTS ii will
display the time adjustment screen (Figure 2.8). An
alarm clock icon is displayed and the time limit set
point is highlighted to indicate the time limit set‐
point in being adjusted
b) Use the increment and decrement buttons to
modify the pressure set‐point value. Momentarily
pressing the button labeled with the up arrow will
increase the set point by 5 min. When the button
is held the set point will increase by 10 min every
second. Momentarily pressing the button labeled
with the down arrow will decrease the set point by
5 min. When the button is held the set point will
decrease by 10 min every second.
c) Press the BACK button to return back to the main
screen or wait until the PTS ii automatically returns
to the main screen.
Time Limit Set Point Adjustment
12
Figure 2.7
Figure 2.8