2.5 
PRESSURE SET POINT ADJUSTMENT 
a)  At the main screen (Figure 2.5), momentarily press 
the  pressure  adjustment  button.  The  PTS  ii  will 
display  the  pressure  adjustment  screen  (Figure 
2.6).    A  pressure  gauge  icon  is  displayed  and  the 
pressure  set  point  is  highlighted  to  indicate  that 
the set point is being adjusted. 
b)  Use  the  increment  and  decrement  buttons  to 
modify the pressure set‐point value.  
 
Momentarily  pressing  the  increment  button  will 
increase  the  set  point  by  5  mmHg.  When  the 
button  is  held  the  set  point  will  increase  by  10 
mmHg  every  second.    Momentarily  pressing  the 
decrement button will decrease the set point by 5 
mmHg. When the button is held the set point will 
decrease by 10 mmHg every second. 
c)  Press the back button to return to the main screen.  
The  PTS  ii  will  automatically  return  to  the  main 
screen after 3 seconds if no buttons are pressed. 
 
Figure 2.6 
Pressure Set Point Adjustment  
d)  Note  to  increase  the  pressure  set  point  above 
475mmHg  confirmation  is  required  (Figure  2.7).  
Momentarily  press  the  OK  button  to  enter 
extended pressure mode or the press back button 
to  return  without  entering  extended  pressure 
mode.    If  confirmed  pressures  above  475  mmHg 
may be selected.  The pressure set‐point value and 
cuff pressure are shown with a yellow background 
when a pressure set point greater than 475 mmHg 
has been selected. 
 
 
Extended Pressure Mode 
2.6 
TIME LIMIT SET POINT ADJUSTMENT 
a)  At the main screen (Figure 2.5) momentarily press 
the  time  limit  adjustment  button.  The  PTS  ii  will 
display the time adjustment screen (Figure 2.8). An 
alarm clock icon is displayed and the time limit set 
point  is  highlighted  to  indicate  the  time  limit  set‐
point in being adjusted 
b)  Use  the  increment  and  decrement  buttons  to 
modify the pressure set‐point value.  Momentarily 
pressing the button labeled with the up arrow will 
increase the set point by 5 min.  When the button 
is held the set point will increase by 10 min every 
second.  Momentarily pressing the button labeled 
with the down arrow will decrease the set point by 
5  min.  When  the  button  is  held  the  set  point  will 
decrease by 10 min every second. 
c)  Press the BACK button to return back to the main 
screen or wait until the PTS ii automatically returns 
to the main screen. 
Time Limit Set Point Adjustment 
12
Figure 2.7 
 
Figure 2.8 
 