CUFF LEAK 
The unit has detected a large 
leak in the cuff, hose or 
connectors which is defined as 
the cuff failing to reach the 
pressure set point in a 
reasonable time,  or the pump 
running excessively while 
regulating the pressure and the 
pressure set point is not being 
adjusted. 
 
 
 
 
CUFF PRESURIZED DURING 
POWER UP  
The unit has detected an 
approximate pressure of 
50mmHg or more in the cuff 
when the unit was powered up.  
The unit assumes a procedure is 
in progress, changes the 
pressure set point to the 
detected pressure, immediately 
regulates the cuff to this 
pressure, and activates alarm to 
alert the user of the condition. 
 
 
 
TOURNIQUET TIME LIMIT 
EXCEEDED 
The unit's elapsed timer, which 
advances when the cuff is 
inflated, has reached the time 
limit set point and caused the 
time alarm to activate. 
 
Alarm message alternates with cuff 
pressure display
Alarm message alternates with cuff 
pressure display
Alarm message alternates with elapsed 
time display
20
 
Connections, hose, and cuff should be 
checked and the leak repaired 
OR 
Reduce pressure set point until unit can 
maintain pressure 
OR 
Press and hold 'DEFLATE' for 2 seconds to 
deflate the cuff
 
 
 
The system assumes that a procedure is 
in progress and adopts the sensed 
pressure as the new set point.  The 
operator should immediately check the 
set value to determine if it needs reset 
Press the pressure adjustment button 
and verify the cuff pressure set point 
 
 
 
The surgeon should be warned of the 
time up condition.  Only on the direction 
of the surgeon, should the time limit be 
set to new value. 
Press the time limit adjustment button 
and increase the time limit. 
OR 
Press and hold 'DEFLATE' for 2 seconds to 
deflate the cuff 
 
 
 
 