OPERATOR AND MAINTENANCE MANUAL 
DELFI PTS ii PORTABLE TOURNIQUET SYSTEM 
U.S. PATENTS 6,213,939; 5,931,853; 5,649,954   PAT. PEND. and FOREIGN PATENTS 
 
 
 
REF  9‐2200‐001X 
Delfi
 
Delfi Medical Innovations Inc. 
Vancouver BC, Canada 
800‐933‐3022 (US & Canada) 
604‐742‐0600 (Global) 
Fax. 604‐742‐3800 
 
www.delfimedical.com 
 
medical
innovations inc.
 
 

Advertisement

Table of Contents
loading

Summary of Contents for Delfi PTS ii

  • Page 1 OPERATOR AND MAINTENANCE MANUAL  DELFI PTS ii PORTABLE TOURNIQUET SYSTEM      REF  9‐2200‐001X  U.S. PATENTS 6,213,939; 5,931,853; 5,649,954   PAT. PEND. and FOREIGN PATENTS      Delfi medical innovations inc.     Delfi Medical Innovations Inc.  Vancouver BC, Canada  800‐933‐3022 (US & Canada)  604‐742‐0600 (Global)  Fax. 604‐742‐3800    www.delfimedical.com   ...
  • Page 2 LIMITED TWO YEAR WARRANTY (North America. only)  SCOPE OF WARRANTY  Delfi Medical Innovations Inc. (‘Delfi’) warrants the  PTS ii Portable Tourniquet System (‘product’) for two years from  date of purchase. During the warranty period, Delfi will repair or replace, at its option, any product which is defective in  materials or workmanship or which fails to meet the published specification for that model. This Limited Warranty is  made only to the original purchaser of the product and is non‐transferable. The remedies described in the Limited  Warranty are the exclusive remedies for breach of warranty. THIS WARRANTY SHALL NOT APPLY TO ANY PRODUCT  WHICH HAS BEEN ALTERED, MODIFIED, DISASSEMBLED OR SERVICED BY ANYONE OTHER THAN DELFI STAFF IN ANY  WAY, OR WHICH HAS BEEN SUBJECTED TO MISUSE OR ABUSE.  DISCLAIMER OF IMPLIED WARRANTIES  The foregoing Express Limited Warranty is given in lieu of any and all other express or implied warranties. DELFI MAKES  NO OTHER WARRANTIES INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR  PURPOSE.  LIMITATION OF REMEDIES  In no case shall Delfi Medical Innovations Inc. be liable for any special incidental or consequential damages whether  based on breach of warranty or other legal theory. Some states do not allow limitations on warranties or on remedies  for breach in certain transactions. In such states, the limits in this paragraph and the preceding paragraph do not apply.  WARRANTY CLAIMS  In the event of a warranty claim within the warranty period please take the following steps:  1. Notify Customer Service Department, Delfi Medical Innovations Inc. at 800‐933‐3022. Please provide details about  the nature of the problem and include the product serial number. Upon receipt of this information, Delfi will provide a  date for service or a return shipping authorization.  2. Upon receipt of the shipping authorization, forward the equipment, freight prepaid, to the location specified in the  shipping authorization.    Your compliance with these steps will help assure that you receive prompt warranty service for your product.        WARRANTY (OUTSIDE North America)  Please contact Delfi for warranty information.          Unit Serial Number _________________    AC Power Supply Serial Number _________________   ...
  • Page 3: Table Of Contents

      TABLE OF CONTENTS  Delfi PTS ii TOURNIQUET SYSTEM    SECTION  TITLE                  PAGE      SECTION 1.0  GENERAL INFORMATION .................... 3     SPECIFICATIONS........................3     INTENDED USE........................4     CONTRAINDICATIONS......................4     PRECAUTIONS IN USE ......................4    ...
  • Page 4: Section 1.0

    Guidance and manufacturer's declaration - electromagnetic emissions Pressure Regulation:  The PTS ii tourniquet system is intended for use in the electromagnetic environment +/‐ 6 mmHg of set point (10‐second average under non  specified below. The user of the PTS ii tourniquet system should assure that is used in such an environment. transient conditions without external leaks)  Emissions Test...
  • Page 5: Intended Use

    1.4  PRECAUTIONS IN USE  1.2  INTENDED USE   The  tourniquet  system  must  be  kept  well  calibrated  The Delfi PTS ii Portable Tourniquet System is intended to  and  in  operable  condition.  Accessories  should  be  be used by qualified medical professionals to temporarily  checked regularly for leaks and other defects.  occlude blood flow in a patient's extremity during surgical   The  tourniquet  cuff  must  never  be  punctured;  procedures  on  that extremity.  Tourniquets are generally  therefore  towel  clips  used  near  the  system  must  be ...
  • Page 6: Adverse Effects

     Whenever  IVRA  (Bier  Block  anesthesia)  is  used,  it  is  guination before tourniquet application may be done  without  the  use  of  an  elastic  bandage  by  elevating  recommended that the tourniquet remain inflated for  the limb for 3 to 5 minutes.   at least 20 minutes from the time of injection.   In  case  of  failure,  the  tourniquet  cuff  must  be  fully  1.5  ADVERSE EFFECTS   deflated  and  the  limb  exsanguinated  again  before  A ...
  • Page 7: Section 2.0

      If  the  unit  is  damaged,  notify  the  carrier  and    Delfi  immediately.    If  the  initial  inspection  results  are  satisfactory,  a  functional and calibration check should be  Icon  Function  performed  after  an  10  hour  charge.    The  attention  label ...
  • Page 8 AC Power Supply  The  red  alarm  indicator  light  is  illuminated  when  any  An AC power supply adapter is supplied with every PTS ii  alarm  condition  exists.    The  alarm  indicator  will  remain  unit.  It is a sealed unit designed specifically for the PTS ii.   illuminated until the alarm condition is corrected, as long  Contact  Delfi  if  your  power  supply  needs  service  or  as sufficient power is available.  replacement.  Plug the connector on the AC power supply  8.  Battery Indicator Symbol  cord into the AC power receptacle on the PTS ii unit (see  The  battery  indicator  symbol  in  the  center  of  the  color ...
  • Page 9: Figure 2.1: Pts Ii Controls, Indicators, Displays, And Connectors

    Figure 2.1: PTS ii Controls, Indicators, Displays, and Connectors                                                                   ...
  • Page 10: Figure 2.2:  Pts Ii Single System Mounting Bracket (7-2200-008)

    Figure 2.2: PTS ii Single system mounting bracket (7‐2200‐008)                                    Figure 2.3: PTS ii Twin system mounting bracket (7‐2200‐009)                                     ...
  • Page 11: Initial Setup

    AC power is connected it remains on constantly.  WARNING:     Use  only  a  Delfi  REF  7‐2200‐020  or  7‐2200‐021  AC  power  supply  and  cord  assembly  Figure 2.2  Welcome Screen  supplied  with  your  PTS  ii.    Do  not  use  any  other  AC ...
  • Page 12 Calibration‐check is displayed (Figure 2.4).  Locking  Connector  to  Luer  adapter  supplied  with  During this time, the unit is checking its calibration  your  PTS  ii  unit  to  connect  the  end  of  the  PTS  ii  at zero pressure.   hose to the reference gauge.    b)  Power  up  the  unit  by  pressing  the  ‘ON/STANDBY’  Figure 2.4  button.  Calibration Check  c)  Enter  the  pressure  adjustment  mode  by  pressing  the pressure adjustment button. ...
  • Page 13: Pressure Set Point Adjustment

    2.5  PRESSURE SET POINT ADJUSTMENT  Figure 2.7  Extended Pressure Mode  a)  At the main screen (Figure 2.5), momentarily press  the  pressure  adjustment  button.  The  PTS  ii  will  display  the  pressure  adjustment  screen  (Figure  2.6).    A  pressure  gauge  icon  is  displayed  and  the  pressure  set  point  is  highlighted  to  indicate  that  the set point is being adjusted.  b)  Use  the  increment  and  decrement  buttons  to  modify the pressure set‐point value.  ...
  • Page 14: Tourniquet Time Reset

    Figure 2.10  2.7.  TOURNIQUET TIME RESET  System Settings and Special Features  The  elapsed  tourniquet  time  may  be  reset  to  zero  at  any time after the cuff has been inflated.  a)  At  the  time  limit  set  point  adjustment  screen  (Figure  2.7)  momentarily  press  the  TIME  RESET  button.  The  reset  confirmation  screen  will  be  shown  (Figure  2.9).    Press  the  OK  button  to  reset  the ...
  • Page 15: Set Default Pressure

    Figure 2.12  2.10  SET DEFAULT PRESSURE  Leak In Progress  The  pressure  set  point  will  be  set  to  the  default  pressure when the system is powered on.  a)  At  the  main  screen  (Figure  2.5)  press  the  settings  button to enter the settings menu (Figure 2.10).   b)  To adjust the default pressure set point, move the  highlighting  box  to  the  pressure  gauge  icon  and  press the OK button.  c)  Use  the  increment  /  decrement  buttons  to  adjust  the ...
  • Page 16: Printer Label Size

    (Figure 2.10).   The  PTS  ii  may  be  used  with  an  optional  industry  standard  thermal label printer (Delfi REF 9‐2200‐150 ).   b)  To adjust the alarm volume, move the highlighting  The  printer  is  supplied  from  Delfi  with  an  appropriate  box to the speaker icon and press the OK button.  interface cable for use with either a single PTS ii system  c)  Use  the  increment  /  decrement  buttons  to  adjust  or with two PTS ii systems in a Twin configuration.   ...
  • Page 17: Operation

      Follow  the  instructions  included  with  the  printer  to  setup  and  connect  the  printer  to  the  PTS  ii  system(s).   Figure 2.21  Printed Event Record (two labels)  Note: When using a printer with two PTS ii systems in a  Twin  configuration  both  PTS  ii  systems  must  be  powered on in order to print.  During  a  surgical  procedure  the  PTS  ii  maintains  a  record ...
  • Page 18 For  each  patient,  tourniquet  pressure  required  to  on  the  patient's  anatomy,  age,  and  absence  of  occlude blood flow to the operative site should be  vascular  disease.    In  many  operating  rooms,  it  is  set  to  the  minimum  effective  pressure.  The  customary  to  prominently  note  the  time  of  minimum  effective  pressure  depends  on  many  inflation, ...
  • Page 19: Alarm Conditions

    is held (in the absence of leaks).  It is also likely that a  2.17  ALARM CONDITIONS  tone  will  sound  under  these  conditions.    The  ‘safe  There  are  a  number  of  conditions  for  which  the  PTS  ii  state’ mode helps prevent unexpected loss of occlusion  will  produce  visual  and  audible  alarms.  Those  during a procedure if a sudden failure occurs.  conditions,  indications  and  appropriate  actions  are  shown in Table 2.1.  The appropriate actions indicated  Although  it  is  very  unlikely,  internal  hardware  failures  are based on the most probable causes and should only ...
  • Page 20 Table 2.1: Alarms and Warnings        Condition Appropriate action / Remarks        TEST FOR LEAKS  This warning is only given when the cuff  is deflated.    The unit has detected leakage  from the cuff, hose, or  If during the time that a cuff is inflated  connectors during the time the  the PTS ii detects minor leakage from the  cuff was inflated.  cuff, hose, or connectors this warning is  given.    Press the OK button to clear the  warning message  Inspect the cuff, hose, and connectors for  damage.  Perform a leak test (see 2.X).                  CUFF PRESSURE HIGH  Normally caused by transient conditions  such as patient movement, regulator  The unit has detected high  overshoot, or hose occlusion.  This  pressure in the cuff.  High  condition, for an extended period, would  pressure is defined as pressure  indicate a hardware failure and the PTS ii ...
  • Page 21         CUFF LEAK  Connections, hose, and cuff should be  checked and the leak repaired  The unit has detected a large  leak in the cuff, hose or  OR  connectors which is defined as  Reduce pressure set point until unit can  the cuff failing to reach the  maintain pressure  pressure set point in a  reasonable time,  or the pump  OR  running excessively while  Press and hold ‘DEFLATE’ for 2 seconds to  regulating the pressure and the  deflate the cuff   pressure set point is not being  adjusted.    Alarm message alternates with cuff      pressure display         CUFF PRESURIZED DURING  The system assumes that a procedure is  POWER UP   in progress and adopts the sensed  pressure as the new set point.  The  The unit has detected an  operator should immediately check the  approximate pressure of ...
  • Page 22       LOW BATTERY  Plug unit in. If the unit is not plugged in, a  battery failure condition will occur and  The battery voltage has dropped  the unit will shut down in a fail safe  to a level indicating that the unit  mode closing all valves.  While running  should be plugged into AC  with a Low Battery Voltage Alarm  power for continued use, and to  Condition other alarm conditions can  charge the battery.  not be guaranteed.  OR  Press ‘ON/STANDBY’ to shut unit down.      Alarm message alternates with elapsed    time display       CUFF NOT FULLY DEFLATED  Disconnect hose to cuff to exhaust  pressure   The unit has detected pressure  in the cuff greater than  15mmHg, 30 seconds after the  cuff has been deflated.      Alarm message alternates with cuff    pressure display    ...
  • Page 23       CHECK PRINTER  Check that the printer is not out of labels;  the labels are correctly installed; and the  The unit has detected that the  printer door is fully closed.   printer is not ready to print.      Alarm message alternates with elapsed  time display           NO PRINTER  The unit was unable to communicate  with the optional printer.  Check to make  The unit can not communicate  sure the printer is correctly connected to  with the optional printer.  the PTS ii, and powered on.            Alarm message alternates with elapsed    time display     BATTERY FAIL  Press ‘ON/STANDBY’ to shut unit down.   When cuff deflation is required,  The battery voltage has dropped  disconnect hose to cuff.  Ensure that the  below the threshold of safe ...
  • Page 24: Section 3.0

    Delfi  staff  poses  a  risk  of  electric  performed as described in this section to ensure continual  shock, damage to the unit, and injury to the patient and  safe and effective operation. This section also serves as a  will void all warranties.  Internal parts in the PTS ii can  guide  to  troubleshooting  and  expediting  unscheduled  only  be  serviced  at  the  factory  by  Delfi  staff.    Please  maintenance.  The maintenance intervals listed below are  contact Delfi if you have problems with your PTS ii unit  provided as a guideline; refer also to the policies in your  that  cannot  be  resolved  by  following  the  maintenance  practice  setting  for  general  tourniquet  maintenance ...
  • Page 25: Periodic Maintenance

    3.2  PERIODIC MAINTENANCE  signal  from  the  pressure  transducers  during  normal  operation.  The calibration factors are stored in memory.  3.2.1  CLEANING  EQUIPMENT REQUIRED:  The  exterior  of  the  unit  may  be  cleaned  as  often  as   Calibrated 0 to 650 mmHg pressure gauge.   required  with  a  cloth  that  has  been  dampened  (not   Adjustable 0 to 650 mmHg pressure source.  dripping)  with  a  mild  detergent.    The  exterior  of  the  ...
  • Page 26: Battery Testing And Replacement

    The  PTS  ii  features  automatic  battery  charging  and  c)  Install  the  new  battery  pack  supplied  by  Delfi  by  monitoring  functions  and  attempts  to  charge  the  aligning  the  keyed  plug  and  pushing  the  connector  battery  whenever  the  unit  is  connected  to  AC  power, ...
  • Page 27: Internal Hardware Servicing

      Internal  parts  in  calibration,  and  troubleshooting  procedures  described  the  PTS  ii  can  only  be  serviced  at  the  factory  by  Delfi  above  do  not  restore  normal  operation,  contact  Delfi  staff.    Please  contact  Delfi  if  you  have  problems  with ...
  • Page 28 Green AC indicator light does not  AC power supply assembly not  Ensure wall socket is working, of the  illuminate when unit is plugged in  plugged in to suitable wall outlet.  correct voltage, and that the plug is all the  to AC power.  way in.  AC power supply assembly not  plugged in to PTS ii unit.  Ensure connector is fully engaged.    Incorrect AC power supply  assembly.  Ensure AC power adapter is the one  supplied with the PTS ii unit.  AC power supply not working  Contact Delfi.  AC indicator light not working.  Contact Delfi.  Red Alarm indicator light does not  Alarm indicator light not working.  Confirm that alarm indicator light  illuminate during alarm conditions.  illuminates during self‐check upon power‐   up.  Contact Delfi.  No cuff pressure and/or tourniquet  Faulty pressure/time display.   Confirm that all segments of the display  time reading.  illuminate during self‐check upon power‐   up.  Internal hardware failure.  Contact Delfi.  Pump runs continuously.  External leak (cuff or hose).  Correct leak to clear leak alarm.  Internal leak. ...
  • Page 29: Figure 3.1: Labels

    ST O RE AT ROO M TEMPERATURE OR CO OLER. REF 7-2200-021 O NLY.   DO NOT INCINERATE. Delfi Me d ic a l Innova tions Inc . REC YC LE / DISPOSE AT PROPER FAC ILITY Delfi Va nc ouver BC Ca na d a 800-933-3022  ...
  • Page 30                         Delfi medical innovations inc.   Delfi Medical Innovations Inc.  Vancouver BC, Canada  800‐933‐3022 (US & Canada)  604‐742‐0600 (Global)  Fax. 604‐742‐3800    www.delfimedical.com          © 2009 Delfi Medical Innovations Inc.  Abatis Medical Technologies Limited Ballybrohan, Killaloe Printed in Canada  0197 County Claire, Ireland (Eire) 4‐2200‐025 Rev. 2 ...