Removing The Lcd; Replacing The Lcd - Fluke 1587 Calibration Manual

Insulation multimeters
Hide thumbs Also See for 1587:
Table of Contents

Advertisement

Removing the LCD

With the PCA removed, the final step in disassembling the Meter is to remove the
LCD assembly from the PCA. Refer to Figure 6 and use the following procedure to
remove the LCD assembly:
1. Remove the two screws from the display end of the PCA.
2. Hold the PCA face down with the fuse-end of the PCA toward you. Locate the
screw-hole near the center of the assembly. This hole is the one used to attach
(with a screw) the PCA to the top case.
3. Now, locate the gray plastic tab just above and to the left of the hole. Using your
thumbnail press the tab down and toward the display end of the PCA. This will
release the LCD assembly from the PCA.
To avoid damaging the plastic guide pins on the LCD
assembly, keep the LCD assembly parallel to the PCA when
separating the two parts.
4. Without tilting the PCA, lift it straight up and away from the LCD assembly.
5. To keep loose parts from falling away from the LCD assembly, keep it face-
down, and set it in a safe place.

Replacing the LCD

The LCD assembly consists of four pieces as shown in Figure 6:
Translucent light-dispersing back panel
Flexible elastomeric conducting strip
Plastic Bezel for housing the assembly's components
Glass LCD display
With the LCD assembly removed from the PCA, use the following procedure to
replace the LCD:
1. Lift the translucent light-dispersing back panel off of the plastic bezel.
2. Remove the elastomeric strip from its slot in the bezel.
Figure 5. Insulation Terminal Clips
W Caution
Insulation Multimeters
Disassembling and Reassembling the Meter
ebp11f.eps
13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1577

Table of Contents