Removing and Reinserting the Circuit Board Assembly

Once the case has been opened, the shields can be disconnected from the circuit board
assembly as follows:
1. Remove the one Phillips-head screw securing the back shield to the circuit assembly.
2. The front shield can now be disconnected from the circuit assembly by detaching the
3. To reinsert the circuit assembly, push the front shield on so that the four clips engage
4. Turn the assembly over, and replace the Phillips-head screw and back shield.
10
Figure 1. Disassembly Details
Then remove the back shield.
four snaps (one at a time) found on the top-front.
Be gentle when detaching or attaching the four snaps. Excessive force can deform or
fracture the snaps.
gently and simultaneously.
Ensure that the shields are tightly attached. Properly fitted shields are required for
the Meter to perform to specification.
ek7f.eps