Do not pump any cleaning solution or other liquids through the vacuum pump, i.e. through the Vacuum Connector. This can damage the LSU. Use only accessories supplied by Laerdal or one of its authorized dealers to help ensure that the LSU operates satisfactorily.
- Remove the LSU from its packaging and inspect all parts for damage. If there are any signs of damage or if any parts are missing, immediately notify Laerdal or one of its authorized dealers. - Insert the battery according to section 5.6.
2.5 Limited Warranty The LSU comes with five (5) year limited warranty*. See the enclosed “Laerdal Global Warranty” for terms and conditions. The warranty is also available at www.laerdal.com * Excluding the canister, tubing systems and battery. LSU DfU US 6791 rev H.indd 5...
User Interface TEST-Button Power ON Indicator Vacuum Indicator External Power Battery Status Indicator Indicator Failure Mode Indicator Operating Knob 3.2 User Interface 3.2.1 Operating Knob The Operating Knob is a combined ON/OFF switch and vacuum selector, located at the bottom of the User Interface. The LSU is switched ON by turning the Operating Knob to the required setting.
Page 8
3.2.2 Indicators 3.2.2.1 Power ON Indicator This green LED has 3 modes: - It is lit continuously while the LSU is switched ON. - It flashes rapidly (approx. twice per second) during the Device Test. - It flashes slowly (approx. once per second) while the Automatic Power-save Function is activated, if the Device Test is interrupted and when the battery is discharged.
Page 9
This green LED bargraph has 3 functions: During operation from internal battery - indicates approx. remaining battery capacity. Note: Immediately after turning the LSU ON and immediately after switching from external power to internal battery operation, all 4 LED’s will flash for 5 seconds before the remaining battery capacity is displayed.
Page 10
LED 1 (lower LED) lit = STEP 1, LED 2 lit = STEP 2 etc. 3.2.3 TEST-Button This button allows you to run a 4-step user initiated Device Test program to identify whether the LSU operates satisfactorily, is assembled correctly or if it needs service: - STEP 1 - Checking for occlusions in the Suction System (including canister and tubing).
To restart the test, set the to “0” and then start over again. Operating Knob 3.3 Bemis Canister ™ The disposable Canister consists of: - A 1200 ml transparent plastic Canister. - Bacterial Filter - 99.99% DOP rated. - Shut-Off Valve and Foam Guard seals tightly to prevent fluid overflow. Canister Lid VACUUM Inlet.
3.4 Power Sources / Connector Sockets The LSU is delivered with an AC Power Cord for connection to AC mains and a DC Power Cord for connection to DC mains. The device can be operated from the internal battery, and can be operated or charged from the following external Power Sources: - AC mains when used with the AC Power Cord: (Max.
4. Opera tion 4.1 Getting Started Note: Read all the Cautions and Warnings listed in section 1 thoroughly before you use the LSU. 1. Check that no parts are missing and that all parts are clean. 2. If you decide to operate the LSU from external power, connect either external AC- or DC-power to the LSU as described in section 3.4.
4.2 Replace the Canister Note: It is recommended to replace the Canister when it is 3/4 full. When liquid reaches the top of the Canister, the LSU will not suction any more liquid. If overflow of liquid from the Canister into the pump is suspected, the LSU must be returned for service (see section 6).
4.3 After Use - When suction is complete, set the Vacuum Switch to “0” and inspect all parts for damage or excessive wear. Replace parts if necessary. - Clean the LSU according to section 5.3. - Perform the Device Test according to section 3.2.3. - Place the LSU on charge (see section 5.4).
5.4 Charge the Battery The internal rechargeable battery can be charged directly from external AC- or DC-power as described in section 3.4. No external charger is needed. To charge the battery, proceed as follows: 1. Make sure the Operating Knob is set to “0”. Connect either external AC- or DC-power to the LSU and charging will start automatically.
5.6 Replace the Battery Note: Use only batteries recommended by Laerdal or one of its authorized dealers. 1. Open the Battery Door. 2. To remove the battery, 3. Withdraw the battery push and move it slightly to from the LSU.* the left and then release.
*If you are discarding the battery, dispose of safely according to local protocols. 7. Trouble Shooting FAULT CONDITION ACTION The LSU does not operate External Power Indicator is Check power cord connec- with the AC or DC Power not lit when the Operating tions and the external AC- or Cord connected.
9. S pec ific ations 9.1 Classification Electrically powered medical suction equipment for field and transport use, according to ISO10079-1. High vacuum/high flow. Not suitable for use in the presence of flammable liquids or gases. Internally powered/class ll equipment type BF, according to IEC 60601-1. Protection class IP34D*, according to IEC 60529.
Page 22
Dismantle the LSU by unscrewing the screws on the back side of the LSU, remove the components and sort according to the table in 9.7. The cleaned LSU can also be delivered to your local Laerdal Medical representative for dismantling/disposal. Laerdal Medical AS will not charge any fee for the dismantling;...
Page 23
9.8 Symbols Direct Current. Alternating Current. Class II Equipment, according to IEC 60601-1. Type BF applied part, according to IEC 60601-1. IP34D Protection class ref. IEC 529 This product is in compliance with the essential requirements of MDD 93/42/EEC as amended by Council directive 2007/47/EC and Council directive 2011/65/EU relating to 0434 restriction on the use of certain Hazardous Substances (RoHS 2).
Page 24
Guidance and manufacturer’s declaration – electromagnetic immunity The Laerdal Suction Unit (LSU) is intended for use in the electromagnetic environment specified below. The customer or the user of the LSU should assure that it is used in such an environment.
Page 25
Guidance and manufacturer’s declaration – electromagnetic immunity The Laerdal Suction Unit (LSU) is intended for use in the electromagnetic environment specified below. The customer or the user of the LSU should assure that it is used in such an environment.
Page 26
Guidance and manufacturer’s declaration – electromagnetic emissions The Laerdal Suction Unit (LSU) is intended for use in the electromagnetic environment specified below. The customer or the user of the LSU should assure that it is used in such an environment.