Getting Started CPRmeter 2 Placement Feedback Display Overview Compression Feedback Depth On Soft Surface Rate Compression Counter Inactivity Debriefing Battery Indicator Maintenance After Each Use Cleaning Disinfection Storing the CPRmeter 2 Between Use Customer Service Indicator Specifications Symbol Glossary Regulatory Information...
® device powered by a replaceable battery. The device is intended for The CPRmeter 2 is not intended for use on SCA victims under 8 years old. use by responders who have been trained in CPR and use of the CPRmeter 2.
Overview Overview Status Light Display Area Membrane Placement guide On/Off Port for adhesive Compression Area Placement guide for adhesive Battery Hatch...
Before Use - Insert Batteries Before Use - Apply Patient Adhesive Cautions • Ensure Patient Adhesives are within their expiration date. • Adhesives should be removed from the device and disposed of after 2 years.
Ensure the patient is on a firm surface. • Remove clothing from the patient’s chest. Turn On Status Light turns green for a few seconds, when CPRmeter 2 is turned on. Place CPRmeter 2 Begin CPR. Provide chest compressions according to your CPR protocol.
CPRmeter 2 Placement Feedback Display Overview Cautions • If the CPRmeter 2 moves during use, re-position it to the center of the chest, as shown. • If difficulty is encountered in applying the device, do not delay initiation of CPR. Remove the device and begin compressions.
Depth On Soft Surface If the CPRmeter 2 detects a compression that exceeds 70 mm (2.75"), it will show the depth indicator below the target area. If a specific CPR event requires CPR to be performed on a patient lying on a...
Inactivity 0:23 Timer Too fast • After 3 seconds the CPRmeter 2 displays an inactivity timer Compression Counter which counts the seconds since the last compression. When compressions are started a counter will display grey up to • After 20 seconds since the last compression, the inactivity timer 25 compressions.
• When the CPRmeter 2 is used in a new CPR event, the Q-CPR Quick Review will display the current event’s statistics. Press the On/Off button once to cycle to the next display.
(do not insert a contaminated CPRmeter 2 into its casing). A large low-battery 2. If it is visibly soiled, wipe the CPRmeter 2 with a soft cloth or icon appears when paper towel to remove as much contamination as possible.
Cleaning Storing the CPRmeter 2 between Use If the CPRmeter 2 has been used in a training situation, it may be Store the CPRmeter 2 in its protective cover to shield the display wiped using an alcohol wipe (70% ethanol solution).
100 to 120/min ± 3/min Material Foam pad with biocompatible adhesive on each side. Shelf Life 2 years when applied to the CPRmeter 2 or 4 years CPRmeter 2 [REF 801-00233] in unopened package. Do not exceed the expiration date on the packaging.
Association. CPRmeter Patient Adhesive The used Patient Adhesive may be These CPRmeter 2 patient adhesives are disposable and are contaminated with body tissue, fluid, for single patient use only. Do not re-use. Re-use will lead to or blood. Dispose of it as infectious increased risk of cross contamination, and/or degradation of waste.
It is not necessary to remove the device from the patient. • Do not practice by using the CPRmeter 2 on a person. It may be used KC symbol (for Korea) with a training manikin or simply on a compliant surface for practice.
Page 17
Regulatory Information Cautions Federal Communications Commission Statement and Industry • Do not apply the CPRmeter 2 to an open wound or recent incision site. Canada Statements • The device is designed to be used only with Laerdal-approved accessories This device complies with part 15 of the FCC rules and RSS-210 of the and may perform improperly if non-approved accessories are used.
Page 18
Electromagnetic use in the electromagnetic environment specified in the tables below. Test Level Level Environment - The user of the CPRmeter 2 should assure that it is used in such an Guidance environment. Electrostatic ±6 kV contact ±6 kV contact...
Page 19
To assess the electromagnetic environment due to fixed RF d=1.2√P transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the CPRmeter 2 is used 0.01 0.12 0.23 exceeds the applicable RF compliance level above, the CPRmeter 2 should be 0.38...
Page 20
Warranty The Laerdal CPRmeter 2 has a one-year limited Warranty. Refer to the Laerdal Global Warranty for terms and conditions. About this edition The information in this applies to the model 801-00233 CPRmeter 2. This information is subject to change.