LW Scientific MX5 Centrifuge Instruction Manual

LW Scientific MX5 Centrifuge Instruction Manual

Advertisement

Quick Links

Buttons

SET:
Change parameters
ENTER:
Enter parameter changes
:
Up
:
Down
RCF / RPM:
Toggle display of Relative Centrifugal Force (g)
/ Speed (rpm)
START:
Start a cycle
SHORT SPIN:
Instantaneous
closed, press the Short Spin button. The machine
will begin a short cycle. The short cycle ends
with the release of Short Spin button.
STOP / OPEN:
Stop a cycle / Open lid
P. 770.270.1394
MX5 Centrifuge
MX5 Centrifuge
centrifugation.
With
the
F. 770.270.2389
Display
SPEED:
Displayed in rpm
TIME:
Time in MIN:SEC
ROTOR:
ID number of rotor: Calculates RCF.
Reference table on last page.
PROG:
Allows recall of 1-9 programs with specific
cycle parameters to include rotor ID
ACC:
Acceleration Rate 1-9, 1 is slowest, 9 is quickest
DEC:
lid
Deceleration Rate 1-9, 1 is slowest, 9 is quickest
865 Marathon Parkway Lawrenceville GA 30046
Instruction Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MX5 Centrifuge and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for LW Scientific MX5 Centrifuge

  • Page 1 MX5 Centrifuge MX5 Centrifuge Instruction Manual Buttons Display SET: SPEED: Change parameters Displayed in rpm ENTER: TIME: Enter parameter changes Time in MIN:SEC ROTOR: ID number of rotor: Calculates RCF. Down Reference table on last page. RCF / RPM: PROG:...
  • Page 2 DO NOT OPERATE THE CENTRIFUGE AT THIS POINT. The lid on the MX5 centrifuge remains locked while at rest and while spinning. To open the lid while at rest, push the STOP/OPEN button, then press on the lid within 3 seconds to release the lid lock. After 3 seconds, the lid will relock. If power fails, the lid can be manually opened by inserting the Manual Lid Release Tool into the release hole on the front right side and left side of the unit.
  • Page 3 If spinning fecals, use the same fecal solution in the balance tube, because water is much lighter than the dense fecal solutions (for Fecal Hints, please contact LW Scientific). Proper balancing will improve sample separation and will extend the life of the centrifuge. Spinning out-of-balance loads may break tubes, and can cause damage to the unit which will not be covered under warranty.
  • Page 4 Carefully press and release on the right side and then left side until the lid opens. This may require gently pushing and lifting on the lid as the release key is inserted. If you have questions, contact LW Scientific service.