With the clamp disconnected from any conductor, press O
1.
to turn the Meter on and press A.
2.
Clamp the jaw around the conductor under test. The Meter
displays the measured conductor current. See Figure 2.
A positive reading indicates current flowing in the
direction of the arrow on the clamp.
A negative reading indicates current flowing in the
opposite direction of the arrow.
Do not clamp more than one wire. Currents cancel and
no results are returned.
The small secondary display shows the reading in terms of
percentage span.