Appendix
Table 2: Troubleshooting Procedures
SYMPTOM
Compressor does not run
34
Installation, Operation and Service Manual
POSSIBLE CAUSE
1.
No power.
2.
Low voltage.
3.
Inoperative control.
4.
Loose wiring at terminals.
5.
Inoperative compressor contactor.
6.
Open or grounded compressor windings.
7.
Defective starting components.
8.
Stuck / seized compressor.
TEST AND CORRECTION
1.
Check power source. If none, call
qualified electrician.
2.
Read supply voltage while unit is
running. Reading must equal rated
±10%.
3.
Refer
to
the
procedure.
4.
Inspect all electrical connections.
5.
Start
compressor
compressor analyzer.
6.
Disconnect compressor leads and
read phase to phase and for any
phase to ground. If electrical checks
are OK, then, check the compressor
start gear.
7.
Replace starting components.
8.
Test
by
using
analyzer, if unsuccessful, replace
compressor.
Thermo Scientific XBF40-MD Series Blast Freezers
controller
setup
using
a
a
compressor