Page 1
LUDLUM MODEL 78 AND 78-1 DIGITAL ANALOG STRETCH SCOPE March 2021 Serial Number 299335 and Succeeding Serial Numbers...
Page 2
LUDLUM MODEL 78 AND 78-1 DIGITAL ANALOG STRETCH SCOPE March 2021 Serial Number 299335 and Succeeding Serial Numbers...
Page 3
RETURN OF GOODS TO MANUFACTURER If equipment needs to be returned to Ludlum Measurements, Inc. for repair or calibration, please send to the address below. All shipments should include documentation containing return shipping address, customer name, telephone number, description of service requested, and all other necessary information.
Environmental Conditions for Normal Use Warning Markings and Symbols Cleaning and Maintenance Precautions Maintenance Recalibration Batteries Technical Theory of Operation Detector Input/Amplifier Discriminator Low Voltage Supply High Voltage Supply Detector Overload Microprocessor Audio S1 (Function) S2-S4 Ludlum Measurements, Inc. March 2021...
Page 9
Calibration General Detector Setup Information Exposure Rate Calibration Detector Overload (OVL) Calibration Recycling Parts List Model 78 Digital Analog Stretch Scope 10-1 Main Circuit Board, Drawing 272 x 591 10-1 Cal/Switch Board, Drawing 272 × 352 10-3 Display Board, Drawing 408 × 259 10-4 Chassis Wiring Diagram, Drawing 272 ×...
3.7 m (12 ft) extendable pole. The two energy- compensated, Geiger-Mueller (GM) detectors measure gamma radiation from 0.1 mR/hr mR/hr to 1000 R/hr (1µSv/h – 10,000mSv/h). The Model 78 measures in R (Roentgen) (Roentgen) units while the 78-1 measures in Sieverts.
Check individual item serial numbers and ensure calibration certificates match. The Model 78 serial number is located on the front panel above the analog meter face. Most Ludlum Measurements, Inc. detectors have a label on the base or body of the detector for model and serial number identification.
The LCD then shows the firmware number in the format “P XX YY.” The “XX” is the firmware number, and the “YY” is the firmware version. (The following figure is for example only, to illustrate location of display.) Note: The Model 78 is not sensitive to neutron radiation. Ludlum Measurements, Inc. Page 2-2 March 2021...
Page 13
FAST Note: The slow response position is normally used when the Model 78 is displaying low numbers that require a more stable display. The fast response is used at high count levels. Depress and release the switch.
Model 78 and 78-1 Technical Manual Section 3 Section Specifications : 4-digit LCD with 1.3 cm (0.5 in.) character height DISPLAY 0.1 to 999.9 mR/hr Model 78 RANGE: 1 µSv/h to 9999 mSv/h Model 78-1 RANGE: : within 10% of true value with connected detector DISPLAY LINEARITY : adjustable from 2-100 mV;...
Page 15
Model 78 and 78-1 Technical Manual Section 3 : unit may be used immediately after the LCD initialization sequence is WARM-UP TIME completed – approximately five seconds after turning on : visual and audible adjustable alarm points ALERT/ALARM : 0.001 to 280 × 10...
Identification of Controls and Functions Display The Model 78 utilizes a four-digit LCD readout with a fixed decimal point. The two smaller digits located in the lower right corner of the display indicate exponential power when in the parameter setup mode. The upper right of the LCD exhibits units and multiplier(s) R/hr, mR/hr, or µR/hr;...
To access the internal circuit boards, unlatch the latches at each end of the Model 78. Carefully separate the top chassis from the bottom cover (referred to as a "can"). The can has the audio speaker (unimorph) with a two- conductor cable attached to the main board.
Model 78 and 78-1 Technical Manual Section 4 : a multi-turn potentiometer (approximately 20 revolutions) sets the HV LIMIT (R027) maximum HV limit with the front-panel control adjusted to the maximum clockwise position. It is adjustable from 1250 to 2400 Vdc.
Page 19
All of the setup parameters are stored in the non-volatile EEPROM, which will retain data even after the Model 78 batteries are removed. Changing parameters and information on switchboard controls, other than those detailed above, are covered in Section 8 of this manual.
Model 78 and 78-1 Technical Manual Section 5 Section Safety Considerations Environmental Conditions for Normal Use Indoor or outdoor use No maximum altitude Temperature range of -20 to 50 °C (-4 to 122 °F); may be certified for operation from -40 to 65 °C (-40 to 150 °F) Maximum relative humidity of less than 95% (non-condensing) Pollution Degree 3 (as defined by IEC 664).
European Union. Cleaning and Maintenance Precautions The Model 78 may be cleaned externally with a damp cloth, using only water as the wetting agent. Do not immerse the instrument in any liquid. Observe the following precautions when cleaning or performing maintenance on the instrument: 1.
Maintenance nstrument maintenance consists of keeping the instrument clean and periodically checking the batteries and the calibration. The Model 78 instrument may be externally cleaned with a damp cloth (using only water as the wetting agent). Do not immerse the instrument in any liquid. Observe the following precautions when cleaning: 1.
Model 78 and 78-1 Technical Manual Section 6 Batteries The batteries should be removed and the battery contacts cleaned of any corrosion at least every three months. If the instrument has been exposed to a very dusty or corrosive atmosphere, more frequent battery servicing should be used.
Model 78 and 78-1 Technical Manual Section 7 Section Technical Theory of Operation Detector Input/Amplifier Refer to the Main Board schematic Negative-going detector pulses are coupled from the detector through C021 to for the following: Amplifier U021. R024 and CR021 protect the input of U021 from inadvertent shorts.
Model 78 and 78-1 Technical Manual Section 7 High Voltage Supply High voltage is developed by blocking oscillator Q241, T141, and C244 and rectified by voltage multiplier CR041-CR043, C041-C043, and C141. High voltage increases as current through R241 increases, with maximum output voltage and Q241 saturated.
Board schematic for the following: S1 is a 16-position binary rotary switch that selects the programmable parameters for the Model 78. The switch selects the parameters using the hexadecimal numbering system via buss lines S2-S4 S2-S4 are pushbutton switches that enter/change the variables for each of the 16 parameters.
: a 16-position rotary switch labeled “0-9” and “ .” This switch FUNCTION Switch selects a parameter setup mode for the Model 78. The selector switch must be set to the “0” position for normal instrument operation. Ludlum Measurements, Inc.
POSITION 6 allows changing the time constant (TC) for RESPONSE TIME the current detector setup. If the response is set to “0,” the Model 78 automatically calculates (for variable mode) the time constant based on Ludlum Measurements, Inc. Page 8-2...
Page 29
Model 78 and 78-1 Technical Manual Section 8 the incoming cps. If a variable of 1-199 is entered for TC, the response time becomes fixed. - Response time is varied in proportion to the incoming count Variable Response rate. The two-position (Fast/Slow) toggle switch selects the maximum time constant (TC) for the variable mode.
Baud Rate 9600 LCD Time Off 5 sec Detector Calibration The Model 78 calibration routine consists of entering detector parameters into memory by way of the switch board and adjusting the controls ( ) for DISC the specific detector operating requirements.
Page 31
XPOSURE ALIBRATION To calibrate the Model 78 to exposure rate after setting the HV and DISC potentiometers, start with the following values for DT (Dead Time, Switch Position 1), CC (Calibration Constant, Switch Position 2), and SC (Series Constant Switch Position 3).
Page 32
Model 78 and 78-1 Technical Manual Section 8 (OVL) C ETECTOR VERLOAD ALIBRATION Note: The detector operating voltage (HV) must be determined and adjusted before the adjustment is performed. If the HV is varied or another detector is substituted, must be readjusted. If the overload feature is not used, adjust the control to the maximum counterclockwise position.
Therefore, Ludlum Measurements, Inc. does not suggest one particular method over another, but simply desires to inform its consumers of the range of recyclable materials present in its products, so that the user will have flexibility in following all local and federal laws.
Model 78 and 78-1 Technical Manual Section 10 Section Parts List Reference Description Part Number Model 78 Digital UNIT Completely Assembled Analog Stretch Scope Model 78 Digital Analog Stretch Scope 48-2832 BOARD Completely Assembled Main Circuit Board, Main Circuit Board...
Need help?
Do you have a question about the 78 and is the answer not in the manual?
Questions and answers