Setting The Cooling Water Parameters - TA Instruments DHR Series Getting Started Manual

Starch pasting cell
Hide thumbs Also See for DHR Series:
Table of Contents

Advertisement

3
Connect the GAS/LIQUID outlet port to the upper port on the Starch Cell jacket, using the 4-mm outer
diameter tubing with the 4-mm to 6-mm adapter provided.
4
Using the 4-mm outer diameter tubing connect the lower port on the Starch Cell jacket to the drain, if
the mains water is the supply, or to the return, if a fluid circulator is used.
5
Connect the EVENT B socket on the rear of the AR Rheometer to the EVENT socket on the CCU using
only the cable.
6
Set the air flow rate to 1 L/min. Note that the reading is taken from the center of the float. To set the
flow rate, it may be necessary to raise the temperature of the Starch Cell using the instrument software
to ensure continuous flow.

Setting the Cooling Water Parameters

For optimum temperature control, set the parameters for the cooling water from within TRIOS Instrument
Control Software.
For more information, see Setting the Cooling Water Parameters for the AR Starch Pasting Cell TRIOS
Help topic.
DHR/AR Series Starch Pasting Cell Getting Started Guide
Page 15