Display, function buttons and basic settings (incubator 51269)
On the left outer corner of the control panel is the connection for the electricity plug. It is important to
push the plug in well, to allow the incubator to function properly. As soon as the plug is connected, the
incubator will turn on.
Besides the display (1 and 2), there are three more buttons which operate the incubator (see "de-
scription of the function buttons"), and two small lamps which glow when the heating is warming up
("work") or adjustments are made ("set").
The display can show the temperature in Celsius or Fahrenheit; this can be set individually.
Display 1 shows the temperature in degrees Celsius.
Display 2 shows the air humidity in per cent.
Explanation of the function buttons (incubator 51269)
The incubator has three buttons which are required for the operation of the incubator. The buttons and
their possible combinations will be explained in the following paragraph. Before operating the incub -
ator, ensure that it is properly plugged into the according area.
The buttons are positioned the following from left to right (green and round) in the centre of the operat-
ing panel: "set", "+" and "–".
1. Set: This button allows to set the basic requirements of the incubator.
Press the button once shortly: adjusting the basic temperature (parameter ID: PP).
By selecting the "+" and "–" button, you can set the desired temperature, pressing the
"set" button again will save the entered value.
The preset temperature is 37.8 
Hold the "set" button: You will go to the menu to calibrate the basic parameters (see
table 3).
By pushing the "set" button, you can go through the individual parameters.