Download Print this page

WilTec 51073 Operation Manual page 10

Mini incubator

Advertisement

meters
Alarm too high
P1
temperature
Heating stop
P2
Alarm too low
P3
temperature
Table 1: Adjustment of the finer parameters via the "set" and "+" button
3. "Set" and "–": Holding both at the same will open the menu for adjusting the turning
options.
The display will show the individual parameter, displayed as number/letter combina-
tion (parameter ID).
With the "+" and "–" buttons, you can choose the various parameters.
If you wish to alter a parameter, push the "set" button, then alter the value up or down-
ward via the "+" and "–" buttons.
Pushing the "set" button again, will save the changes.
Turning para-
Parameter
meters
Turning
fre-
F1
quency
Turning
dura-
F2
tion
Table 2: Adjusting the turning via the "set" and "–" buttons
4.
"+" and "-": Holding both pushed at the same time for a longer duration (approx. 8 sec)
resets the incubator back to factory settings, there will be a peep sound
© by WilTec Wildanger Technik GmbH
http://www.WilTec.de
http://www.aoyue.eu
http://www.teichtip.de
ID
range
0–99.9 ℃
0–99.9 ℃
0–99.9 ℃
Adjustment range
ID
0–999 min
0–999 sec
Item 51073, 51269
setting
38.8 ℃
37.8 ℃
Please note that the radiator will
switch
entered value, but will still radiate
some warmth, and the temperature
might still rise a little bit in the incub-
ator. E. g. this means that a temper-
ature of 38 
when setting the temperature to
37.8 
. If you wish to avoid this, the
temperature of the heating stop
needs to be corrected downward.
36 ℃
Standard
setting
90 min
10 sec
off
when
reaching
can be reached,
Note
When adjusting to 0, no
turning takes place.
When adjusting to 0, no
turning takes place.
Page 10
12
the
2020-1

Advertisement

loading

This manual is also suitable for:

51269