Technical Specifications - Vicks VNT275US Owner's Manual

Noncontact infrared body thermometer
Table of Contents

Advertisement

Available languages

Available languages

Technical Specifications

Measuring Distance
Memory
Power Supply
Requirements
Batteries
Operable Voltage Range
Environmental
Operating Environment
Transport and Storage
Environment
Dimensions and Weight
Weight (without batteries) 3.2 ounces (90g)
Size
This infrared thermometer meets requirements established in ASTM Standard E 1965-98 (for the thermometer
system). Full responsibility for the conformance of the product to the standard is assumed by Kaz USA, Inc.,
a Helen of Troy Company, Health & Home, 400 Donald Lynch Blvd., Suite 300, Marlborough, MA 01752
ASTM laboratory accuracy requirements for the thermometer only in the display range of 36 °C to 39°C (96.8
°F to 102.2 °F) for infrared thermometers is ±0.3 °C (±0.5 °F), whereas for mercury-in-glass and electronic
thermometers, the requirement per ASTM Standards E 667-86 and E 1112-86 is ±0.1 °C (±0.2 °F).
This device conforms to the following standards:
IEC 60601-1: Medical electrical equipment. Part 1: General requirements for basic safety and essential
performance
ASTM E1965-98: Standard Specification for Infrared Thermometers for Intermittent Determination of
Patient Temperature
ISO 80601-2-56: Medical electrical equipment - Requirements for basic safety and essential performance of
clinical thermometers for body temperature measurement
IEC 60601-1-2: Medical electrical equipment – Part 1-2: General requirements for basic safety and
essential performance – Collateral Standard: Electromagnetic disturbances – Requirement and tests
NOTE: DO NOT use this device in the presence of electromagnetic or other interference outside the normal
range specified in IEC 60601-1-2.
IEC 60601-1-11: Medical electrical equipment – Part 1-11: General requirements for basic safety and
essential performance – Collateral standard: Requirements for medical electrical equipment and medical
electrical systems used in the home healthcare environment
ISO 15223-1: Medical devices - Symbols to be used with medical device labels, labeling and information to
be supplied - Part 1: General requirements"
6
0.4 inches to 2 inches (1 cm to 5 cm)
50
1.5V (AAA) Alkaline battery x 2 (IEC Type LR03)
2.6V~3.6V
Temperature: 59°F – 95°F (15°C – 35°C), Relative
Humidity≤93% non-condensing, Atmospheric Pressure:
0.7 atm to 1.04 atm (70 kPa to 106 kPa)
Temperature: -4 °F – 131°F (-20°C – -55°C), Relative
Humidity ≤93%, non-condensing Atmospheric Pressure:
0.7 atm – 1.04 atm (70 Kpa – 106 Kpa)
L: 5.4" x W: 3.7" x H: 1.6" (L:138mm X W:95mm X H:40mm)