Product Specifications - Vicks VNT200US Instruction Manual

No touch 3-in-1 thermometer
Hide thumbs Also See for VNT200US:
Table of Contents

Advertisement

Product Specifications

Type:
Measuring range:
Resolution:
Laboratory Accuracy:
This thermometer displays a calculated oral equivalent estimate.
Display:
Acoustic:
Operating temperature:
Automatic Switch-off :
Weight:
Long term storage ranges
Storage/transport temperature:
Humidity:
Battery:
measurements
Pressure:
This infrared thermometer meets requirements established in ASTM Standard E
1965-98 (for the thermometer system). Full responsibility for the conformance of
the product to the standard is assumed by Kaz, USA, Inc., a Helen of Troy Company,
Health & Home, 400 Donald Lynch Blvd., Suite 300, Marlborough, MA 01752.
ASTM laboratory accuracy requirements for the thermometer only in the display
range of 36 °C to 39°C (96.8 °F to 102.2 °F) for infrared thermometers is ±0.3
°C (±0.5 °F), whereas for mercury-in-glass and electronic thermometers, the
requirement per ASTM Standards E 667-86 and E 1112-86 is ±0.1 °C (±0.2 °F).
10
No Touch Thermometer (VNT200)
Body: 34.0 °C – 43.0 °C (93.2 °F – 109.4 °F)
Object: 0 °C – 100.0 °C (32.0 °F – 212.0 °F)
0.1 °C (0.1 °F)
Body Mode:
±0.2 °C for the range 35.0 °C – 42.0 °C
(±0.4 °F for 95.0 °F – 107.6 °F)
±0.3 °C (±0.5 °F) for the range outside
35.0 °C – 42.0 °C (95.0 °F – 107.6 °F)
Surface Mode:
±2.0 °C for the range 0.0 °C – 100.0 °C
(±3.6 °F for 32.0 °F – 212.0 °F)
Liquid Crystal Display, 4 digits plus special
icons
Normal: Green backlight, One long beep for
normal temperature
Elevated or high temperature: Yellow or Red
backlight, 4 double beeps
Erase memory: one long beep
15 °C – 40 °C (59.0 °F to 104 °F)
Approx. 20 seconds after any button is pressed
85.6 g (with batteries), 63.0 g (w/o batteries)
-25 °C to 55 °C (-13 °F – 131 °F)
15–95% non-condensing
(2) AAA Batteries - at least 1000
700-1060 hPA (0.7-1.06 atm)
This device conforms to the following standards:
• EN 60601-1:2014 – Medical Electrical Equipment – Part 1: General requirements
for basic safety and essential performance
• ISO 80601-2-56:2017 – Medical electrical equipment – Part 2-56: Particular
requirements for basic safety and essential performance of clinical thermometers
for body temperature measurement
• ASTM E1965-98:2016 – Standard Specification for Infrared Thermometers for
Intermittent Determination of Patient Temperature
• IEC 60601-1-2:2014 Medical electrical equipment – Part 1-2: General requirements
for basic safety and essential performance – Collateral Standard: Electromagnetic
disturbances – Requirement and tests
NOTE: DO NOT use this device in the presence of electromagnetic or other
interference outside the normal range specified in IEC 60601-1-2:2014.
• ISO 15223-1:2016 - Medical devices - Symbols to be used with medical device
labels, labeling and information to be supplied - Part 1: General requirements
• EN 1041:2013 Information supplied by the manufacturer of medical devices
• IEC 60601-1-11:2015 Medical electrical equipment – Part 1-11: General
requirements for basic safety and essential performance
– Collateral standard: Requirements for medical electrical equipment and medical
electrical systems used in the home healthcare environment
To protect the environment, dispose of empty batteries at appropriate
collection sites according to national or local regulations.
Keep out of reach of young children, elderly and pets.
BF Type applied part.
See Instruction of use.
Protected against solid foreign objects of 12.5 mm diameter and larger.
Protected against water drops falling vertically when the device is tilted
up to 15°.
Helen of Troy cares about the environment, please recycle.