Setting The Commissioning Parameters - Siemens SIMATIC IOT2000 SG-Shield Operating Instructions Manual

Iiot weighing solutions
Hide thumbs Also See for SIMATIC IOT2000 SG-Shield:
Table of Contents

Advertisement

Requirement
You have installed the Node-Red example program for SIMATIC IOT200 SG-Shield.
You have opened the Node-Red dashboard.
Procedure
Calibration with calibration weights
1. Parameterize the calibration weights.
2. Start the "Set calibration point 0" command with the scale empty.
3. Load the scale with the specified calibration weight.
4. Start the "Set calibration point 1" command.
Automatic calibration
1. Specify the load cell parameters (characteristic value and rated load).
2. Ensure that the scale is empty.
3. Start the "Run automatic calibration" command.
6.4

Setting the commissioning parameters

Parameter
Sensor manufacturer
Sensor ordering number
Location designation of sensor
Channel active
Sensor type
0: 4-wire load cell
1: 6-wire load cell
Sampling rate (Hz)
50/60 Hz toggling
0: Disabled
1: 50 Hz
2: 60 Hz
Low-pass filter cut-off frequency
(Page 23)
Ordinal number low pass filter
(Page 23)
Mean value filter depth
(Page 23)
SIMATIC IOT2000 SG-Shield
Operating Instructions, 08/2021, A5E50985404-AA
Format
Length
[bytes]
string24
24
string24
24
string24
24
unsigned16
2
unsigned16
2
float
4
unsigned16
2
float
4
unsigned16
2
float
4
6.4 Setting the commissioning parameters
r/w
Default
r/w
0
r/w
0
r/w
0
r/w
1
r/w
0
r/w
100.00
r/w
1
r/w
2.0
r/w
4
r/w
0.1
Parameter assignment
Min
Max
Mod‐
bus ad‐
dress
-
-
1000
-
-
1012
-
-
1024
0
1
1073
0
1
1077
1.00
120.00
1081
0
2
1089
0.05
50.0
1093
1
10
1101
0.01
3.413333 1105
19

Advertisement

Table of Contents
loading

Table of Contents