Advertisement

4. Operation

4.8. Re-programming pH measurement

Default programming (standard)
Resolution
• 0.01 pH
Measuring mode
• By stability .
- Standard criterion (variation < 0 .01 pH in 6 s) .
• In continuous .
- The result is stored when the measurement
has finished .
- The result is printed when the measurement
has finished .
View mV
• No
Limits
• No
pH isopotencial
• pHiso 7.00
"Re-programming" flow chart
10-04-2010
17:15
pH MEASURE
New CAL in 12 h
Channel 1
...by stability
Measuring Conditions
Resolution: 0.01
...
Meas.: By stability
Start Measurement
View mV
Channel 1
Limits
Isopotential: pH 7
Start measurement
Measuring Conditions
Resolution: 0.01
Meas.: By stability
View mV
Limits
Isopotential: pH 7
Measuring Conditions
Resolution: 0.01
Meas.: By stability
View mV
Limits
Isopotential: pH 7
Measuring Conditions
Resolution: 0.01
Meas.: By stability
View mV
Limits
Isopotential: pH 7
Measuring Conditions
Resolution: 0.01
Meas.: By stability
Start of
View mV
Limits
measurement with
Isopotential: pH 7
new programming.
etc.
Other options
• 0.1 pH
• By stability.
- Rapid criterion (variation < 0 .02 pH in 6 s) .
- Standard criterion (variation < 0 .01 pH in 6 s) .
- Strict criterion (variation < 0 .002 pH in 6 s) .
• In continuous . Selection of the time interval for:
- Data storage .
- Data printing .
• By time . Selection of the time interval for:
- Data storage .
- Data printing .
• Simultaneous display of pH and mV concerned.
• Higher and lower limits. Acoustic warning when the measure-
ment is out of limits, in the printed ticket appears "A" beside
the measured value .
• pHiso 7.00
Important note:
Once the measurement has been re-programmed, it will be performed in
accordance with the new parameters .
It is highly recommended to apply the same stability criteria to measure and
calibrate .
Measuring Conditions
Resolution pH:
0.1
0.01
0.001
Measuring Conditions
Measuring Conditions
Meas.:
By Stability
By stability
Criteria
In Continuous
Rapid
By time
Standard
Strict
Measuring Conditions
Measuring Conditions
Meas.:
In Continuous
By stability
In Continuous
Acquis. Interval
By time
01:00
Measuring Conditions
Measuring Conditions
Meas.:
By time
By stability
In Continuous
Measurement time
By time
00:01:00
Measuring Conditions
View mV:
Yes
No
Measuring Conditions
Measuring Conditions
Limits:
Limits
Yes
Upper: pH 8.20
No
Lower: pH 6.05
Measuring Conditions
Isopotencial:
pH 7
See Electrode isopotential on page 19.
Data introduction
Calculate
• 0.01 pH
• 0.001 pH
• Manual introduction
• Value calculation.
Measuring Conditions
By Stability
Criteria
Rapid
Standard
Strict
Measuring Conditions
In Continuous
Acquis. Interval
Data storage in
01:30
the Data Logger. It
coincides with the
printing and PC
Measuring Conditions
sending. Selecting
By time
0 store/print at
Acquis. Interval
the end of the
00:01:00
measurement.
Measuring Conditions
Upper limit:
pH
8.20
11