Auxiliary D/A Output
Range
Resolution
Environmental
Operating Temperature
Specification Temperature
Potentiostat Mode
Applied E Range
Accuracy
DC Bias
Scan Ranges
Resolution
Drift
Noise and Ripple
Galvanostat Mode
Applied i range
DC accuracy
Scan Ranges
Current Interrupt
Measurement Type
Cell Switching Time
Minimum Interrupt Time
Maximum Interrupt Time
A/D converter
Resolution
Accuracy
Timing
General
Power
Leakage i (floating, earthed Working Electrode)
Appendix B -- PCI4/750 Specifications --
± 5 volts or 0 to 10 volts
2.5 mV
0-70 °C (inside computer)
25 °C
± 11 volts
± 2 mV ± 0.3% of setting
± 8 V
± 6.4 V, ± 1.6V, and ± 0.4 V
200 µV/bit, 50 µV/bit, 12.5 µV/bit
< 30 µV/C
< 20 µV rms (1Hz - 10 kHz)
± full scale current (no 3 nA range)
± 0.3% full scale
± 2X full scale current
(sample 2 points on decay, extrapolate)
< 1 µsec (1 kΩ cell )
15 µsec
64 msec
16 bits
0.1% of full scale
50 µsec to 600 sec
45 W maximum
< 1 A at 12 volts
< 0.08 A at -5 V
< 1 nA @ DC
6-4
< 6.5 A at +5 V
< 0.02 A at -12 V