Appendix A -- PCI4/300 Specifications
Control Amplifier
Compliance Voltage
Output Current
Unity Gain Bandwidth (software selectable)
Slew Rate (software selectable)
Differential Electrometer
Input Impedance
Input Current
Bandwidth (-3dB)
CMRR
Voltage Measurement
Full Scale Ranges
Resolution(16 Bits)
DC Accuracy
Offset Range
Current Measurement
Analog Full Scale Ranges
Controller Board Gains
Resolution (16 bits)
Offset Range
DC Accuracy (with 1X Controller Board Gain)
Bandwidth (-3 dB)
Auxiliary A/D Input (see Appendix E)
Range
Bandwidth
Input Impedance
NOTES:
1. All specifications subject to change without notice
2. Offset specifications apply after software calibration
Appendix A -- PCI4/300 Specifications --
> ± 20 volts @ 150 mA
> ± 300 mA
>1 MHz, > 200 kHz, > 90 kHz, > 20 kHz
>50 V/µsec, > 10 V/µsec, > 5 V/µsec, > 1 V/µsec
12
> 10
Ω in parallel with 5 pF
< 10 pA
> 4 MHz
> 100 dB (DC to 2 kHz), >60 dB @ 100 kHz
± 30V (±12 V usable), ± 3V, ± 300 mV, ± 30 mV
1 mV/bit, 100 µV/bit, 10 µV/bit,, 1 µV/bit
± 0.3% Range ± 1mV
± 12 V with 1.5 mV resolution
± 3 nA to ± 300 mA in decades
1, 10, 100
0.1 pA/bit to 10 µA/bit
± 4X full scale (only 2X full scale is useful)
± 0.3% range ± 50 pA
> 500 kHz (300 µA—300 mA full scale)
> 100 kHz (30 µA full scale)
> 10 Hz (3nA full scale)
± 3 volts differential
20 Hz
100 kΩ