SEFRAM MW 3031 User Manual page 25

Ac/dc 400a ac digital clamp meter
Table of Contents

Advertisement

Available languages

Available languages

sine wave at full scale and non-sine wave
below half scale. For non-sine wave add the
following Crest Factor corrections :
For Crest Factor of 1.4 to 2.0, add 1.0% to
accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to
accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to
accuracy.
Max.CF of 400.0V range : 2.5
Max.CF of 600V range : 1.7
Input Impedance : 10MΩ, <100pF.
Overload protection : 600V rms.
DCV
Range
400.0V
600V
Input Impedance :
Approx. 10MΩ, <100pF.
Overload protection : 600V rms.
Resistance / Continuity
Range
400.0Ω
4.000kΩ
40.00kΩ
400.0kΩ
4.000MΩ
40.00MΩ *
* There is a little rolling less than
Open Circuit Voltage : Approx. 0.4V
Resolution
0.1V
1V
Resolution
0.1Ω
0.001kΩ
0.01kΩ
0.1kΩ
0.001MΩ
0.01MΩ
25
Accuracy
± (0.6%+ 2 D)
± (0.7%+ 5 D)
Accuracy
± (0.9%+ 5 D)
± (0.9%+ 2 D)
± (1.5%+ 5 D)
± 2%

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mw3035