6.3. Diagnostic Mode Readings

In Diagnostic mode, you can step through readings by pressing [N/-].
6.3.1.

Sensor Raw Count

Sensor Raw Count is indicated by "rAU" followed by three dashes (- - -). This indicates that the
sensor has digital output that does not produce a raw count.
• Press [N/-] to advance to the next reading.
• Press [MODE] and [Y/+] together for three seconds to exit Diagnostic Mode and enter
Detection Mode.
6.3.2.

Battery Raw Count

Battery Raw Count is indicated by "bAt" followed by a number.
• Press [N/-] to advance to the next reading.
• Press [MODE] and [Y/+] together for three seconds to exit Diagnostic Mode and enter
Detection Mode.
6.3.3.

Temperature Raw Count

Temperature Raw Count is indicated by "tNp" followed by a number.
• Press [N/-] to advance to the next reading.
• Press [MODE] and [Y/+] together for three seconds to exit Diagnostic Mode and enter
Detection Mode.
6.3.4.

Zero Raw Count

Zero Raw Count is indicated by "0rC" followed by three dashes
(- - -). This indicates that the sensor has digital output that does not produce a raw count.
• Press [N/-] to advance to the next reading.
• Press [MODE] and [Y/+] together for three seconds to exit Diagnostic Mode and enter
Detection Mode.
6.3.5.

Calibration Delta Counts

Calibration Delta CTS is indicated by "CdC" by three dashes (- - -). This indicates that the sensor
has digital output that does not produce a raw count.
• Press [N/-] to return to the first raw count.
• Press [MODE] and [Y/+] together for three seconds to exit Diagnostic Mode and enter
Detection Mode.