INPUT: 120 VAC, 2 A, 1 phase, 60 Hz OUTPUT: 0-35,000 VAC Due to the short circuit current limiting of the 635-5P, the following maximum full output voltages are available at the specified currents. Output Current Approx. Max. Voltage Available...
INTRODUCTION The Phenix Model 635-5P is a portable AC Dielectric Test Set for performing dielectric strength tests primarily on vacuum bottles. The test set may also be used for testing other samples which are no load or light load (< 1 mA) in nature. A built-in self-protection feature current limits the short circuit output current to safe levels, preventing damage to the test set.
SAFETY Because the 635-5P is designed to supply high voltages, safeguards have been built into it. These safeguards are only effective if safe work practices are also used. The test set and these procedures have been designed for use by knowledgeable electrical workers who have been properly trained in its use.
CONTROLS AND INSTRUMENTATION The following is a brief description of the various controls and instrumentation found on the Model 635-5P. Refer to Figure 1. 1. Main Power On Switch. Toggle switch supplies power to test set. 2. ON Indicator Lamp. Neon lamp indicates Main Power is ON.
SET-UP AND OPERATION 1. Place unit on bench or other suitable flat surface. 2. Undo (2) latches on side of case. 3. Remove cover, power, control, return, and output cables. Exercise caution in removing cover because control cables are permanently connected to the bottom of the case. 4.
ELECTRICAL SCHEMATICS Drawing Number Description 1. -01-E634 Model 635-5P Portable AC Dielectric 2. -01-P118 Dual 5 VDC Power Supply PC Board 3. -01-P119 Meter Calibration PC Board...
MAINTENANCE No solution or chemical stronger than ordinary mild soap and water solution should be applied to the cabinet area of this unit. Care must be used when cleaning the meter faces and console panel. Abrasives may remove printing and descriptive titles and scratch meter faces. When cleaning, always have unit disconnected from power source.
18 months from the date of shipment. The obligation of Phenix Technologies, Inc. under this warranty is limited, in its exclusive option, to repair, replace, or issue credit for parts or materials which prove to be defective, and is subject to purchaser's compliance with the Phenix Technologies, Inc.
Page 12
Phenix Technologies, Inc. and any determination so made shall be final and binding. This warranty is expressly in lieu of all other warranties expressed or implied on the part of Phenix Technologies, Inc., including the warranties or merchantability and fitness for use, and consequential damages arising from any breach thereof and Phenix Technologies, Inc.
Serial Number Reason for Return Cause of Defect If Phenix Technologies, Inc. deems return of the part appropriate, it will then issue an "Authorization for Return". If return is not deemed advisable, other inspection arrangements will be made. NOTE: Material received at this plant without the proper authorization shall be held as "Customer's Property"...
When your purchase order is received at our office, a representative of Phenix Technologies will contact you to confirm the current price of the part being ordered. If a part you order has been replaced with a new or improved part, an Applications Engineer will contact you concerning any change in part number.
12-1 RECOMMENDED SPARE PARTS In order to maintain your set in full operating condition with a minimum of down time, the following spare parts should be kept on hand to avoid unnecessary phone calls, expensive modes of shipment, delays in repairs, etc. Pricing is available upon request. Quantity Description Part Number...