Center The Fine Tuning; Disable Scaling - Honeywell NK User Manual

Portable indicator
Table of Contents

Advertisement

3.4 Center the Fine Tuning

Center the
and
fine-tuning potentiometers to allow precise adjustment
zero
span
in either direction. Use a small screwdriver to turn each potentiometer up to 15
turns clockwise until it clicks. Then, turn it back halfway, 7-½ turns
counterclockwise (see Figure 4: on page 17).

3.5 Disable Scaling

Scaling is the ratio of the output voltage to the displayed reading. Disable scaling
so it does not interfere with calibration. Use a small screwdriver to turn the
scaling
potentiometer up to 15 turns clockwise until it clicks (see Figure 4: on page 17).
Calibration
16