Honeywell NK User Manual page 27

Portable indicator
Table of Contents

Advertisement

4.
—This control scales the display to the desired units. Determine
Scaling
the scaled shunt-cal value using the following formula:
With no load on the transducer, press and hold the
small screwdriver to adjust the
reads the scaled shunt-cal engineering units.
After scaling, return the panel to the case and tighten the four captive fasteners. If
the display does not read 0000, fine tune the unit again according to the procedure
on page 12.
Calibration
potentiometer until the display
scaling
button. Use a
shunt cal
23