Support Routine maintenance must be performed on the Auralis system before every use or every week (if for long term use) to maintain safety and reliability of the equipment. See section Care and preventive maintenance for more information. If you require further information, please contact Arjo for comprehensive support and maintenance to maximize the long-term safety, reliability and value of the product.
For the Auralis Seat Cushion, the patient weight must not exceed 200 kg (440 lb). • For the Auralis Plus mattress, the patient is within the weight range of 100 kg (220 lb) to 454 kg (1000 lb). • For the Auralis Plus mattress, the patient Body Mass Index (BMI) is >30...
Safety instructions WARNING WARNING To avoid serious injur y or death, the To avoid severe burn injur y, do not use the Auralis pump in the presence of and accessible at all times. WARNING To avoid strangulation, always use WARNING...
IFU. When the Auralis system is in use, make sure: • The Auralis system is kept away from sources of heat and moisture, and protected from dust, lint and dirt. • The Auralis system is not covered.
The Auralis Plus mattress is designed to be used with the Arjo Citadel Plus bed frame. The Auralis mattress range (narrow, standard and Plus) may also be used with other bed frames (Arjo or non- Arjo). An assessment should be made by clinician or caregiver to determine which mattress and bed frame combination to use.
Actions before every use (2 steps) 2. If any item is damaged - do NOT use the product. After each patient The position of the caregiver The caregiver should be positioned in front of the pump during operation. See Figure 1 Figure 1 HOME CARE...
8. Non-slip base with: Foam underlay (mattress 175) 9. Securing straps (four on each side) 10. Overlay base (including micro air loss system) 11. Overlay zip 12. Tube-set 13. Connector 14. Cable management PARTS DESIGNATION - AURALIS 175 AND 200 MATTRESS REPLACEMENT...
Run mode - press the button once. The indicator Figure 2 Standby mode - press the button for two seconds.The NOTE To turn the Auralis pump off during power failure or battery operation, press the Run/Standby button for four seconds. Wait indicator...
Page 18
The Reactive CLP mode (see Figure 6) maintains a constant lower pressure in all the cells. When activated the indicator is shown. The Auralis seat cushion does not have Reactive CLP mode. Comfort control buttons and indicators The Comfort control buttons (see Figure 7) set the comfort level by adjusting the pressure within the support surface.
Page 19
1 hour battery life left. The alarm can be muted (see Mute alarm button). When there is less than 15 minutes of battery life left, the Auralis pump enters the Reactive CLP mode. At this point, the audible alarm cannot be muted.
Alarm indicators The power and alarm indicators are located on the top (the control panel) and front of the Auralis pump. The alarm indicators shown on the front of the pump are green when the pump is operating normally and yellow when there is an alarm condition.
The Heelguard™ cells help to keep the pressure in the heel area lower for longer. Figure 18 Cell hooks The cell hooks attach the cells to the Auralis mattress base. Tube-set The tube-set is anti-kink and not removable from the Auralis mattresses.
Figure 20 Auralis 110 mattress overlay The Auralis 110 mattress is an overlay and is placed on top of an existing bed mattress. Fixing straps (Auralis 110 mattress overlay) four corners of an existing bed mattress and prevent the Auralis 110 mattress overlay from sliding off.
This allows the mattress replacements to be attached to different types of bed frames. In addition, Auralis to avoid mattress movement due to the increased patient weight.
The detachable cover is attached by a zip. Alternating cells The seat cushion cells provide alternating support to the patient. The Auralis seat cushion can only be used with Alternating mode. Fixing straps from falling off a chair with an open construction or removable seat cushion.
Page 25
In the event of the patient suffering a cardiac activated: buttons simultaneously. See Figure 25 unit is open. See Figure 26 Figure 25 the main body of the mattress in 15 seconds. unit unit until it clicks into place. See Figure 27 Figure 26 Figure 27 CPR RAPID DEFLATION UNIT...
Auralis mattress with the protective Figure 29 top cover. Place the detachable cover over the Auralis mattress overlay with the Arjo logo visible and located at the foot end of the Auralis mattress. Continue with steps on the next page. ASSEMBLE THE AURALIS SYSTEM...
5. Zip the cover onto the Auralis mattress overlay, start from the head end. Do not trap any material in the zip. Make sure the zip is fully closed. closed position. (See Figure 30.) The CPR rapid Auralis 175, 200 and Plus mattress...
(Auralis Plus mattress only.) See Figure 34. Auralis pump (7 steps) Figure 34 1. Hang the Auralis pump at the foot end of the bed. See Figure 35 Make sure the pump is not: • near a heat source •...
5. Place the cable in the cable management. studs. See Figure 37 7. Connect the power cable to a power source. Auralis Plus mattress extensions (accessory only) (5 steps) The mattress extensions are intended to be used when the bed frame is operated in fully extended length mode.
Check that there are no sharp objects Figure 40 on the chair which may puncture the seat cushion. Place the Auralis seat cushion on top of the chair seat. From a position of standing and facing the front of the chair, make sure: •...
Page 31
Place the protective cover over the Auralis seat cushion. Make sure the Arjo logo is visible at the front part of the seat. 4. Zip the cover onto the Auralis seat cushion. Do not trap any material in the zip. Make sure the zip is fully closed.
The Auralis Plus mattress takes approximately 17 2. When the Auralis head end of the bed. 3. Place the patient on the Auralis mattress with the head at the head end and the feet at the foot end. See Figure 46...
Page 33
360 kg (793 lb). (Auralis Plus only) If there is a power failure, the Auralis pump enters Battery mode. If the Auralis pump is fully charged the Auralis pump offers a minimum of 3 hours operational use.
Extend (1 step) • mode by 5 minutes. times. For further extension repeat step 1. If not extended, the Auralis pump goes back to the previous mode by default. Deactivate of the other modes. Alternating mode, default setting Activate (1 step) •...
The Reactive CLP mode indicator is shown. Deactivate Figure 50 To deactivate Reactive CLP mode, select either therapy is required, remove patient from the surface and turn off the Auralis pump. Lock out mode Locking • Push and hold Comfort control Positive (+) and Comfort control Negative (–) buttons at the same...
To avoid entanglement when the Auralis pump is in Transport mode, make sure that the tube-set connector is stowed appropriately. NOTE There is no Transport mode on the Auralis seat Figure 53 cushion. Deactivate To deactivate the Transport mode, connect the tube- set.
Page 37
through narrow openings. 1. Locate the valve symbol on the bottom of the mattress. See Figure 55 2. Turn the valve (under the valve symbol) on the bolster fully clockwise. See Figure 56 3. Repeat on the other bolster, if required. NOTE Figure 55 bed frame into its narrowest form.
1. Press the Run/Standby button for 2 seconds. The Auralis pump enters Standby mode. 2. Disconnect the power cable from the power source. 3. To turn off the Auralis pump completely, press the Run/Standby button for 4 seconds until a double beep is heard. Figure 58...
1. Clean and disinfect the Auralis pump according to page 40. 2. Wrap the power cable around the hanging brackets. See Figure 62 3. Store the Auralis pump in a designated area, in Figure 62 TURN OFF AND STORE THE AURALIS SYSTEM...
Cleaning and disinfection instructions The Auralis system should be routinely cleaned and disinfected after each patient and at regular intervals while in use, as local practice for all reusable medical devices. If there are any questions regarding cleaning and disinfecting the equipment contact Arjo Customer Service.
Auralis pump. 9. Remove any excess moisture by wiping over with a dry cloth. 10. Allow the Auralis pump to air dry before reuse. Clean the Auralis mattress / seat cushion 11. Spray all external areas of the Auralis mattress/...
Page 42
13. Use water and a new cloth to wipe off all traces of cleaning solution, followed by a dry cloth to remove any excess moisture. 14. If it is necessary to clean the inside of the Auralis mattress/ Auralis seat cushion, unzip the top cover to access the internal components.
Page 43
27. Allow the internal components to air dry before reattaching the top cover of the mattress. 28. If required, the straps and base of the Auralis mattress/seat cushion can also be cleaned using the same method as above. Unclip the straps before cleaning for ease of access.
Expected service life The Auralis pump has an expected service life of 7 years. To maintain the condition of the Auralis pump, the Auralis pump must be serviced regularly according to the schedule recommended by Arjo. CARE AND PREVENTIVE MAINTENANCE...
Page 45
Visually check all cell hooks Auralis system Visually check all labels Perform a full functionality test on the Auralis system Auralis Plus mattress extension (accessory only) Visually check the top cover Visually check the foam core Visually check all labels...
See Figure 63 Perform a full functionality test on the Auralis system (6 steps) 1. Connect the tube-set to the Auralis pump. Make sure the tube-set clicks into place. 2. Connect the power cable to a power source. The Auralis pump sounds with a start-up tone and a self-diagnostic check runs for 10 seconds.
Visually check all labels Check that all labels are attached on the Auralis system according to section Labels on page 56 and Labels on Auralis mattresses, Auralis seat cushion and Auralis mattress extensions (accessory only) Labels on Auralis mattresses, Auralis seat cushion and Auralis mattress extensions on page 58.
3. Check the valves on the mattress is not closed. seat cushion. WAIT 4. Check that the tube-set 3. The valves on the Auralis connector is correctly seat cushion are not connected to the Auralis closed. pump - an audible click 4.
Page 49
1. The Auralis seat cushion 1. Disconnect the Auralis seat Not applicable does not is in use. cushion. activate only applies to the Auralis 2. The Auralis mattress is in mattresses. use. button for 2 seconds. Continue on the next page. TROUBLESHOOTING AND ALARMS...
Page 50
The top and bottom Comfort simultaneously. Pump is now unlocked. OVER The Auralis pump is operating 1. Turn off the Auralis pump. 10 seconds TEMPERATURE outside its temperature range 2. Disconnect the power cable. and has entered the transport 3.
Page 51
Intentionally left blank TROUBLESHOOTING AND ALARMS...
Page 52
If the Auralis pump is stored in conditions outside the operating ranges, allow time for its temperature to stabilise to normal before use. Allow a minimum of 8 hours if the Auralis pump is stored at -20 °C or 60 °C.
Page 53
Temperature (Ambient) 0 °C to 40 °C (32 °F to 104 °F) Relative humidity range 0 % to 95 % (non-condensing) CAUTION To avoid damage to the Auralis system: • Do not store the Auralis system in direct sunlight. •...
Page 54
Chlorine concentrations may vary from 250 ppm to 10,000 ppm depending on local policy and contamination status. If an alternative disinfectant is selected from the wide variety available, Arjo recommend that suitability TECHNICAL SPECIFICATIONS...
Page 55
STANDARDS/CERTIFICATE DESCRIPTION IEC 60601-1 3.1 edition Medical Electrical Equipment – Part 1 : General requirements for basic safety and essential performance. IEC 60601-1-11:2015 Medical Electrical Equipment – Part 1-11 : General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment IEC 60601-1-2:2014 (4th EMC Edition)
Labels Labels on the Auralis Pump LABEL EXPLANATION LABEL EXPLANATION Product label Product label States technical performance and requirements, e.g. input power and States technical performance and requirements, e.g. input power and input voltage. input voltage. Serial number label Serial number label...
Page 57
SYMBOL EXPLANATION Read the Instructions for use before use. CE marking indicating conformity with European Community harmonised Indicates the product is a Medical Device according to EU Medical Device Regulation 2017/745. accordance with CAN/CSA-C22.2 No.60601.1 (2008) + (2014) and ANSI/ AAMI ES60601-1 (2005)+AMD(2012).
Page 58
Labels on Auralis mattresses, Auralis seat cushion and Auralis mattress extensions LABEL EXPLANATION Serial number label Top cover label Care and cleaning label States the washing conditions Law tag (US and Canada only) Example of mattress Law tag, if applicable...
Page 59
(lb) Product weight (Auralis mattresses) = kg (lb) Product weight (Auralis seat cushion) Recommended wash temperature: 15 min at 60 °C (140 °F) Max wash temperature: 15 min at 95 °C (203 °F) Tumble dry at 80 °C (176 °F) Maximum drying temperature: 80 °C (176 °F)
Electromagnetic Compatibility (EMC) the user of the Auralis system should assure that it is used in such an environment. Electromagnetic Compatibility (EMC) This product complies with the requirements of applicable EMC Standards. Medical electrical equipment needs special precautions regarding EMC and needs to be installed in accordance with the following instructions: •...
Page 61
GUIDANCE AND MANUFACTURER’S DECLARATION: ELECTROMAGNETIC EMISSIONS (EMI) of the Auralis pump should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment - guidance RF emissions Group 1 The Auralis pump uses RF energy only for its internal function.
Page 62
GUIDANCE AND MANUFACTURER’S DECLARATION: ELECTROMAGNETIC IMMUNITY The Auralis the user of the Auralis pump should assure that it is used in such an environment. Immunity test IEC 60601 test Compliance level Electromagentic environment - guidance level Conducted RF 3Vrms 6Vrms ISM,...
Page 63
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND THE PUMP The Auralis pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Auralis pump can help prevent electromagnetic interference...
Page 64
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY the user of the Auralis pump should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance Electrostatic ± 8 kV contact ±...
Parts and accessories Auralis Plus Long mattress extension Accessory Auralis Plus Short mattress extension Accessory 636614 636615 Coiled Skin IQ power cable 636377 PARTS AND ACCESSORIES...
Page 66
Intentionally left blank PARTS AND ACCESSORIES...
Page 67
AUSTRALIA FRANCE POLSKA Arjo Australia Arjo SAS Arjo Polska Sp. z o.o. Building B, Level 3 2 Avenue Alcide de Gasperi ul. Ks Piotra Wawrzyniaka 2 11 Talavera Road CS 70133 PL-62-052 KOMORNIKI (Pozna ) Macquarie Park, NSW, 2113, FR-59436 RONCQ CEDEX...
Page 68
At Arjo, we are committed to improving the everyday lives of people affected by reduced mobility and age-related health challenges. With products and solutions that ensure ergonomic patient handling, personal hygiene, disinfection, diagnostics, and the effective prevention of pressure ulcers and venous thromboembolism, we help professionals across care environments to continually raise...
Need help?
Do you have a question about the Auralis and is the answer not in the manual?
Questions and answers