display 'A'). If you continue pressing the "MEMORY"button
the last individual results in each case are displayed
with date and time. The average of all stored results and
also the stored individual results are compared with the
currently set Doctor's function limit values.
• By pressing the "MODE"button, you can call up your set
Doctor's function limit values in order to compare.
• note: If you change your limit values during the
measurement, this affects the interpretation of the results
already stored.
After pressing the button "MEMORY" and selecting the
user, you can clear the memory and the Doctor's function
limits by holding down the "MODE" and
5 seconds.

10 | troubleshooting

In case of faults, the "
display. Error messages can occur when
• the inflation pressure is higher than 300 mmHg;
• the blood-pressure value is unusually high or low;
• you move or talk during the measurement (in addition to
"
", the cardiac arrhythmia icon
display);
• inflation takes longer than 16 seconds;
In the above cases, you must repeat the measurement.
Make sure that you do not move or talk. Re-insert the
batteries if necessary, or else replace them.
"up" buttons for
" message appears in the
appears in the
11 | cleaning & care
• Clean your blood pressure computer carefully only with a
slightly moistened cloth.
• Do not use detergents or solvents.
• On no account must you immerse the computer in water,
otherwise liquid can enter it and cause damage.
• Always store the blood pressure monitor in the
storagecase.

12 | technical data

Type
Measuring method
Measuring range
Precision of pressure systolic ± 3 mmHg/
reading
Measuring uncertainty Max. admissible standard
Doctor's function
Memory
Dimensions
Weight
Cuff size
34
BC18
Oscillating, non-invasive blood
pressure measurement on the wrist
pressure 0 - 300 mmHg/
heart rate 40 - 199 beats/min
diastolic: ± 3 mmHg/ pulse:
± 5% of the displayed value
deviation according to clinical
testing: systolic: 8 mmHg/
diastolic: 8 mmHg
sys. 70 - 250 mmHg
dia. 40 - 120 mmHg
3 x 30 storable data lines
L 74 x W 84 x H 33 cm
Approx. 140 g
for a wrist circumference of
approx. 13,5 to 19,5 cm