Pressure Relief Mechanism
Operational at:
Change in maximum recommended
flow when back pressure of 7 kPa
is applied:
Filters:
Safety System:
Width:
Height:
Depth:
Weight:
Shipping Weight:
Operating Ambient Temperature
and Humidity:
Cabinet:
No AP/APG:
1195054-B
241 kPa ± 34.5 kPa (35 psi ± 5 psi)
0.7 L/min
Cabinet (2), Outlet HEPA and Compressor Inlet
Current overload or line surge shutdown.
High temperature compressor shutdown.
Internal High Temperature shutdown.
High Pressure Alarm with compressor shutdown.
Low Pressure Alarm with compressor shutdown.
Battery Free Power Loss Alarm.
SensO
Oxygen System.
2
Possible Obstruction Alert.
46.7 cm ± 1 cm (18 3/8 in ± 3/8 in)
67.0 cm ± 1 cm (26 3/8 in ± 3/8 in)
36.5 cm ± 1 cm (14 3/8 in ± 3/8 in)
24.0 kg ± 1 kg (53 lb ± 2 lb)
27.2 kg ± 1 kg (60 lb ± 2 lb)
10°C to 35°C (50°F to 95°F) at up to 60% relative humidity
Impact Resistant flame-retardant plastic cabinet that conforms to UL 94 V-0
Not suitable for use in the presence of a flammable anaesthetic mixture.
Technical Data