Invacare® Platinum® Oxygen Concentrator
Humidifier Bottle
with Cap
1. Remove cap A from bottle B.
2. Fill humidifier bottle with boiled tap water or bottled
water to the level indicated by the manufacturer. Boil
tap water for approximately ten minutes and cool to
room temperature prior to use.
CAUTION!
Risk of Damage
– Confirm the humidifier cap is not
cross-threaded on the humidifier bottle.
20
Humidifier Bottle without Cap
3. Insert a flathead screwdriver in the plate groove C on
the top edge of the filter access door D and gently pry
the filter access door off.
4. Pull up and remove the humidifier bottle adapter E
(next to inlet filter).
WARNING!
Risk of Injury
To avoid impacting the flow of oxygen:
– Install the humidifier bottle and/or other
accessories properly.