Calibration Check Kit Installation (Optional) - Bosch ACS 653 Repair Instruction

Table of Contents

Advertisement

Introduction
1
1
1
Figure 1-16. Front plastic cover removal
7. Remove the 2 screws (3) and remove the
flush port manifold.
8. Fix the N2H2 manifold by mean of the
2 screws (3).
3
3
Figure 1-17. N2H2 Manifold installation
9. Disconnect rilsan hose from fitting (4) of the
injection manifold and connect it to fitting (6)
of N2H2 manifold.
10. Connect the rilsan hose coming with the kit
within fitting (5) of N2H2 manifold and fitting
(4) of the injection manifold.
11. Connect faston S19 and S20 of solenoids
harness to the solenoids on N2H2 manifold.
12. Insert the rear cover so that the 4 holes on
the sides the cover fit into the 4 clips on the
machine. Push down to hook the rear cover.
SP00D00624
2021-04-08
S20
6
2
2
Figure 1-18. N2H2 Manifold hoses connection
13. Close the service rear door.
14. Reinstall the plastic frontal cover and
fix it by 6 screws (1) and the 2 ones (2)
(see Figure 1-16).
15. Attach the outside high- (red) and low-side
(blue) service hoses. Tighten at 7.9 Nm torque.
16. Switch on the machine.
17. Call up the Main Menu.
18. Select NEXT icon.
19. Select MENU icon.
20. Select Maintenance
Nx Leak Test and Turn ON.
21. Perform a functional test and check for leaks.
Calibration check kit installation
(optional)
For the magnet installation, refer to Figure 1-19.
1. Plug the magnet (1) within the housing
underneath the machine base.
For fix the calibration weight, refer to
"Calibration check" in this section.
1
Figure 1-19. Calibration check kit installation
24
ACS 653, ACS 663 Service Manual
S19
5
4
Service menu
"
Robert Bosch GmbH
"

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Acs 653-pAcs 663Acs 663-pAcs 663-ri

Table of Contents