Emc Compliance - Omega HHP400 User Manual

Smart manometer and sensors omega tethered sensors suite application
Hide thumbs Also See for HHP400:
Table of Contents

Advertisement

M20011 REL-0000 (2103)

EMC compliance

Conformity to EN 61326-1:2016
This product has been evaluated to EN 61326-1:2016 the EMC
General requirements for electrical equipment for measurement,
control and laboratory use. This evaluation demonstrated that
when the product is exposed to Radiated Radio Frequency
energy in the frequency range 130-1000 MHz and 1.6-2.1 GHz, the
pressure and temperature readings may fluctuate beyond the
published tolerance of each respective value being measured. In
the event of such a phenomenon occurring, the operator should
move away from (or have removed) any source of strong RF
emissions such as radio transmitters, cell-phones, etc., and repeat
the measurement, or reconnect to a PC.
Electrostatic discharge (ESD)
This unit is sensitive to ESD discharges that may result in the
unit re-starting of the operating system, or interruption of a PC
connection. In the unlikely event of such an occurrence, users
should ensure they discharge any static charge they may be
carrying by touching a known grounded conductive surface prior
to operating the equipment .
User Manual for Smart Manometer
59 of 60

Advertisement

Table of Contents
loading

Table of Contents