Download Print this page

ResMed VPAP ST Information Manual page 6

Positive airway pressure device
Hide thumbs Also See for VPAP ST:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
 
Aircraft use
Electromagnetic
compatibility
IEC 60601-1
classification
Device technical specifications
Pressure and
flow state
Mode reading
Physical
Air filter
6
ResMed confirms that the device/s meets the Federal Aviation Administration
(FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air
travel.
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
environments.
It is recommended that mobile communication devices are kept at least 1 m
away from the device.
Information regarding the electromagnetic emissions and immunity of this
ResMed device can be found on www.resmed.com, on the Products page
under Service and Support. Click on the PDF file for your language.
Class II (double insulation), Type BF, Ingress protection IP21
Operating pressure range (measured at the mask): 4-20 cm H
25 cm H
O (VPAP)
2
Maximum single fault steady state pressure: 30 cm H
for > 6 sec; 40 cm H
Pressure measurement tolerance: ±0.5 cm H
Flow measurement tolerance: ±6 L/min or 10% of reading, whichever is greater,
at 0 to 150 L/min positive flow
CPAP mode
Set Pressure: 4 to 20 cm H
S, ST and T modes
IPAP: 4 to 25 cm H
O; EPAP: 2 to 25 cm H
2
Nominal dimensions (L x W x H): 153 mm x 140 mm x 86 mm
Weight: 835 g
Housing construction: Flame retardant engineering thermoplastic
Air outlet: 22 mm conical air outlet (complies with ISO 5356-1:2004)
Hypoallergenic air filter: Acrylic and polypropylene fibers in a polypropylene
carrier
Standard air filter: Polyester non-woven fiber
O - if pressure exceeded for >1 sec
2
O
2
O (CPAP), 2-
2
O - if pressure exceeded
2
O (± 4% of the measured reading)
2
O
2

Advertisement

loading

This manual is also suitable for:

Vpap sS9 vpap stS9 vpap s