SECTION 7 PARAMETERS & CALIBRATION

7.1 Enter setting mode

The scale has 10 parameters that can be set by the user plus a method of entering
the calibration section.
To set parameters or calibration, hold TARE/UNIT key ,then press ON/OFF key ,untill
it shows F0 CAL .

7.2 Calibrate

After enter setting mode, the display will show the first function,
"F0 CAL". press
TARE/UNIT key, display shows "UNLOAD", empty the platter, after STABLE indicator
on, press TARE/UNIT key again, display shows the calibrate weight, you can press
ZERO key to select calibrate weight, after you select, press TARE/UNIT key, display
shows "LOAD", put the calibrate weight on the pan , after STABLE indicator on, press
TARE key to sure, display shows "PASS", after BS/BSI selfchecking again, and is
ready for you.

7.3 Change division

After enter setting mode, the display shows "F0 CAL", press ZERO key, display
shows "F1 INC", press TARE/UNIT key to enter, press ZERO key to select
division(range 0.5/1/2), press TARE/UNIT key to sure, press ZERO key and
TARE/UNIT key at the same time to escape.

7.4 Check internal counts

After enter setting mode, display shows "F0 CAL", press ZERO key twice, display
shows "F2 INP", press TARE/UNIT key to enter, display shows internal counts, press
TARE/UNIT key and ZERO key at the same time to escape.

7.5 Set backlight mode

After enter setting mode, display shows "F0 CAL", press ZERO key three times,
display shows "F3 BK", press TARE/UNIT key to enter, press ZERO key change
backlight mode(AU: automatic, OFF: off, ON: on), press TARE/UNIT key and ZERO
key at the same time to escape.

7.6 Set unit

After enter setting mode, display shows "F0 CAL", press ZERO key four times,
display shows "F4 unt" , press ZERO key to change the second Unit.(g: gram, TL:
Taiwanese liang, lb: ounce)

7.7 Set sleep time

After enter setting mode, display shows "F0 CAL", press ZERO key five times,
display shows "F5 OFF", press ZERO key to change the time for sleep .(range
0,3,5,10,20 minute)