Air Compressor

The typical preventive maintenance interval for the air compressor is
15,000 hours. Rebuild the air compressor, if applicable, after 15,000
hours. If the air compressor for the unit you purchased does not come
with a rebuild kit, replace the air compressor if the maximum cycle
pressure starts to go below 179 kPa (26 psig).
Topaz/Topaz+/Topaz Ultra Instruction Manual
7-3