AirSep
Corporation
Circuit Board
The circuit board controls the cycle time and sequence of each solenoid
valve.
Terminal Strip
The terminal strip provides a connection point for the wiring to the circuit
board and the air compressor.
Mixing Tank
The mixing tank provides stable flow and purity for short-term surge in
oxygen demand.
Pressure Regulator
The pressure regulator controls the delivery pressure of the product
oxygen. This regulator is factory set and should never be adjusted.
Exhaust Muffler
The exhaust muffler allows waste gas to exit the adsorbers.
Hour Meter
The hour meter indicates the total number of hours that the unit cycles.
Topaz/Topaz+/Topaz Ultra Instruction Manual
Changing the regulator setpoint in any way can
potentially cause damage to the molecular sieve and
will void the warranty.