Oxygen Pressure Gauge
This gauge indicates the delivery pressure of the product oxygen.
Flowmeter
The flowmeter allows you to view and adjust the flow rate of the product
oxygen. The flowmeter should never be adjusted beyond the setpoint
designated for the unit in Appendix A.
Oxygen Outlet
The 'B' size adapter at the oxygen outlet connects to your application.
Air Exhaust
Heat from the unit escapes through the air exhaust, located on the left
side of the enclosure.
Power Cord
The power cord (not shown) and its grounded plug connect from a
connection (See Figure 4.1) on the unit to a properly grounded electrical
outlet to supply electrical power to the unit.
l/O Power Switch
The l/O power switch starts and stops the operation of the unit. When you
set the switch to l, the green indicator light illuminates.
Circuit Breaker
Use the circuit breaker button to reset the unit after an electrical overload
shutdown.
Fuse
For the 240 V units, an additional fuse safety is provided as per the CE
regulations.
4-2
Increasing the flowmeter beyond the setpoint
designated for the particular model can potentially
cause damage to the molecular sieve and will void the
warranty.