2.Pre-Measurement Procedures - 2.2.Turning the Power On/Off
2.2.4.

Setup Date and Time

The current time is displayed on the top right of the screen, but if it
differs greatly from the actual time, the time can be adjusted.
See
2.2.5.

Preparing for More Precise Measurements

Warm up the product for about 60 minutes after turning on the power in order to perform more precise
measurements.
After the warm up is complete, perform
See
2.2.6.

Turning Off the Power

Step 1.
When the
Power
while the power is on, the shut down process starts and the
[Shut down] dialog box indicated below is displayed on the
center of the screen. Tap the
product. Tap the
the power.
If the
Power
dialog box is displayed, the product automatically shuts down.
Make sure to shut down the product to turn off the power, as directly removing the power cable from
the power outlet without shutting down can damage the files in the internal storage.
"8.2.5. Environment
Setup".
"Zero
"4Setup
Measurement". This completes the preparations for measurement.
key on the operation panel is pressed
OK
Cancel
key to continue without turning off
key is pressed again while the [Shut down]
adjust" of the input modules.
key to shut down the
CAUTION