Digital sphygmomanometer

Advertisement

Instruction Manual
Manuel d'instructions
Manual de instrucciones
Manuale di Istruzioni
Bedienungsanleitung
使用手册
WM+PD4001301

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the UM-101 and is the answer not in the manual?

Questions and answers

Summary of Contents for A&D UM-101

  • Page 1 Instruction Manual Manuel d’instructions Manual de instrucciones Manuale di Istruzioni Bedienungsanleitung 使用手册 WM+PD4001301...
  • Page 2: Table Of Contents

    UM-101 Digital Sphygmomanometer Instruction Manual Contents Dear Customers ................2 Preliminary Remarks .................3 Notes for Proper Use.................4 Parts Identification ................6 Symbols.....................7 Using the Monitor ................8 Measurements...................9 Useful Features ................10 Troubleshooting................11 Maintenance ..................11 Technical Data .................12 English...
  • Page 3: Dear Customers

    1. Dear Customers Congratulations on purchasing a state-of-the-art A&D sphygmomanometer, one of the most advanced monitors available on the market today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen. We recommend that you read through this manual carefully before using this device for the first time.
  • Page 4: Preliminary Remarks

    2. Preliminary Remarks Compliance Compliance with European Directive 93/42 EEC for Medical Products The device conforms to the following requirements: European Directive 93/42 EEC for Medical Products; Medical Products Act; European Standards for Electrical Medical Equipment EN 6060-1 (General Safety Provisions), EN 60601-2-30 (Particular Requirements for the Safety of Automatic Cycling Indirect Blood Pressure Monitoring Equipment), EN 60601-1-2 and EN 55011 (Electromagnetic Compatibility);...
  • Page 5: Notes For Proper Use

    Training The healthcare provider should stop the examination if there is an abnormality, such as a patient feeling excessive arm pain, and remove the cuff to protect the patient. Repair Do not attempt to open the device. Contact your nearest A&D authorized representative and they will repair or replace the device.
  • Page 6: After Use

    During Use This device should be used by trained professionals. Stop the examination immediately, if the patient feels pain during a measurement or if the device does not work properly. Stop using the device if you notice any abnormalities (for example; liquid inside the device) and request a full service.
  • Page 7: Parts Identification

    4. Parts Identification Main Body and Accessories Cuff connector Pressure bar display Numerical display MARKER button ON/OFF button Exhaust valve Rubber bulb Cuff 1.5V Batteries Band for attaching a Instruction manual (LR6 or AA) stethoscope on the arm Function ON/OFF Turns on or off the device.
  • Page 8: Symbols

    5. Symbols Symbols Function / Meaning Recommended Action Standby and Turn the device on. Pressure value holding and lifetime MARKER counter. Battery installation guide Serial number Date of manufacture Type BF: Device, cuff and tubing are designed to provide special protection against electrical shocks.
  • Page 9: Using The Monitor

    (Refer to "Attaching the Arm Cuff" in next section) Arm Size Recommended Cuff Size Order No. 33 cm to 45 cm Large cuff AX-UM-101-03 23 cm to 33 cm Medium cuff AX-UM-101-01 17 cm to 25 cm Small cuff...
  • Page 10: Measurements

    Attaching the Arm Cuff 1. Wrap the cuff around the upper arm, about RANGE 2 ~ 3 cm above the elbow as shown on INDEX right. Place the cuff directly against the skin, as clothing may cause a faint pulse and could result in a measurement error.
  • Page 11: Useful Features

    7. Press the button again to turn off the power. ON/OFF Note: Model UM-101 has an automatic power-off function. 8. Useful Features Measurement with MARKER Button You can put a marker at a certain pressure value Marker...
  • Page 12: Troubleshooting

    9. Troubleshooting Problem Possible Reason Recommended Action Nothing Batteries are empty. Replace all batteries with new ones. appears in the Reinstall the batteries with their Battery polarities display, even negative and positive ends matching are not in the when the power those indicated in the battery correct position.
  • Page 13: Technical Data

    11. Technical Data Model UM-101 Measurement method Stethoscopy with stethoscope Measurement range Numerical display Pressure: 0 ~ 300 mmHg Pulse: 30 ~ 200 beats / minute Pressure bar display Pressure: 20 ~ 280 mmHg Measurement accuracy Numerical display Pressure: ±3 mmHg Pulse: ±5 %...
  • Page 14 A&D Co.,Ltd. 1-243 Asahi, Kitamoto-shi, Saitama 364-8585 JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119 A&D INSTRUMENTS LTD. <Authorized Representative Established in the European Community> Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxon OX14 1DY United Kingdom Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485 A&D MERCURY PTY.