Operating environment

W A R N I N G
The instrument may not be used in dangerous environments.
The manufacturer and its suppliers reject any express or indirect guarantee for
use with high-risk activities.
N O T I C E
Protect the instrument from temperature extremes, including heaters, direct
sunlight and other heat sources.
Note the following points so that the instrument will operate perfectly and
have a long service life.
Place the instrument firmLy on an even surface. Do not push any
objects under the instrument.
The ambient temperature must be +15 to +40 °C (+60 to +104 °F).
The relative humidity should be less than 80%; moisture should not
condense on the instrument.
A minimum clearance of 15 cm (5.9 in.) must be ensured above and
on all sides of the device; this allows the air to circulate and
prevents overheating of the electric parts. The dosage arm of the
stepper motor must be able to move freely.
Do not operate or store the instrument in extremely dusty, humid or
wet locations.
Make sure that no liquids enter the instrument and immediately wipe
away any liquids that contact the instrument.
Protect the instrument against vibrations, direct sunlight, corrosive
gases as well as strong magnetic and/or electric fields.
Only use genuine replacement parts and accessories.
Install the instrument in a levelled place and on a compact surface
to minimize the sound level of the instrument. This will avoid
undesired resonances that can increase the sound level.
English
7
Keep the surface of the instrument and all accessories clean and
dry at all times. Splashes or spills on the instrument should be
cleaned up immediately.
The instrument must be serviced by the manufacturer's service
department at the prescribed intervals.
Installation
Installation of the burette
The burette dispenses the titrant: EDTA 0.010 M TitraVer (P/N 20553)
solution.
The burette comprises the syringe, stepper motor and valve. The
stepper motor and valve are already installed in the TitraLab; the syringe
has to be installed by the user.
W A R N I N G
Risk of injury to pinch points and cuts
Make sure that the titrator is never used/operated without the syringe cover.
Do not introduce your fingers on the bottom of the cover while the instrument is
connected.
Make sure that the system is only assembled by qualified and trained personnel.
N O T I C E
Warning against system damage
Make sure that the syringe is never operated without the attachment sleeve.
Only use genuine replacement parts and accessories.
The instrument must be serviced by the manufacturer's service department at the
prescribed intervals.
Note: Make sure that you do not misplace the attachment sleeve for the
syringe. It is not provided with a new syringe.
1.
Push the syringe piston into the syringe body.
Slide the attachment sleeve fully onto the syringe so that the
narrow end points toward the piston.
2.
Remove the syringe cover by unscrewing it.