problem - Problem
2
1
3
2
3
2
1
TST / 11.11.97
possible
causes
No oil or cleaning liquid is coming
- Transport seals are not removed
- U-shaped seals 02687200 dry
- Pump (no.3) defective
- Defective tubing
- Empty tank
No oil in the air motor
- Speed regulator of air motor
in " 0" -position
!
Ball of function indicator steadily
visible
- Air bubble(s) below the ball
- Stucked because of dirt
Ball of function indicator does
not appear
- Pump (no. 3) defective
1
- Tubing defective
- Tubing not completely filled
with liquid yet
- Stucked because of dirt
- Indicator leaky
Ball of function indicator not visible
during operation; visible at standstill
- Air in tubing system
- Possible entry of air at a
defective tube
Instrument does not keep holding
on the adapter
- Not completely sticked on
- Engaging hook defective
W&H Dentalwerk Bürmoos Ges.m.b.H., A - 5111 Bürmoos / Austria
repair procedure
(no.)
9
10
5/6/11
6
12
13
14
14
5/6/11
18/6
14
14
6
14
6
15
6
mögliche
Problembehebung
Ursache
Öl oder Reinigungslösung werden
nicht gefördert
- Transportdichtungen nicht entfernt
- Nutringdichtungen 02687200
trocken
- Pumpe (Nr. 3) defekt
- Leitung defekt
- Tank leer
Kein Öl im Luftmotor
- Drehzahlregler des Luftmotors
auf " 0" -Stellung
Kugel der Funktionsanzeige
ständig sichtbar
- Luftblase(n) unter der Kugel
- Steckt durch Verschmutzung
Kugel der Funktionsanzeige
erscheint nicht
- Pumpe (Nr. 3) defekt
- Leitung defekt
- Leitung noch nicht vollständig
mit Flüssigkeit gefüllt
- Steckt durch Verschmutzung
- Anzeige undicht
Kugel der Funktionsanzeige während
Laufzyklus weg; im Stillstand sichtbar
- Luft im Leitungssystem
- Möglicher Lufteintritt in defekter
Leitung
Instrument hält nicht am
Motoransatz
- Nicht ganz aufgesteckt
- Rasthaken defekt
(Nr.)
9
10
5/6/11
6
12
13
14
14
5/6/11
18/6
14
14
6
14
6
15
6
2 / 6