Instructions for Use
301 plus

Advertisement

Table of Contents
loading

Summary of Contents for W&H assistina 301 plus

  • Page 1 Instructions for Use 301 plus...
  • Page 2: Table Of Contents

    Contents Symbols .......................................... 4 In the instructions for use ..................................4 On the medical device ..................................5 On the packaging/cartridge ................................. 6 1. Introduction ....................................... 7 2. Unpacking ........................................9 3. Scope of delivery ..................................... 10 4. Safety notes ......................................11 5.
  • Page 3 10. Maintenance ......................................28 O-rings replacement ................................... 29 Filter replacement ....................................30 Air filter replacement ..................................31 11. Servicing ........................................ 32 Returns ........................................ 32 Preparation of returns ..................................33 12. Technical data ......................................34 13. Disposal........................................35 Explanation of warranty terms ................................... 36 Authorised W&H service partners ................................
  • Page 4: Symbols

    Symbols In the instructions for use WARNING! ATTENTION! ATTENTION! (risk of injury) (to prevent damage occurring) General explanations with no danger to persons or property CE mark Do not dispose with domestic waste WARNING! Flammable liquid and vapour...
  • Page 5: On The Medical Device

    Symbols On the medical device Consult instructions for use DataMatrix Code Catalogue number for product information including UDI (Unique Serial number Date of manufacture Device Identification) Manufacturer CE mark Medical device...
  • Page 6: On The Packaging/Cartridge

    Symbols On the packaging/cartridge This way up Permitted temperature Keep away from heat range for storage Fragile, handle with care CE mark Use by Medical device WARNING! Keep dry Flammable liquid and vapour EUH210 safety data sheet “Der Grüne Punkt” (The Green Consult instructions for use available on request Dot) trademark of Duales...
  • Page 7: Introduction

    1. Introduction For your safety and the safety of your patients Prior to initial use please read the Instructions for use. These explain how to use your medical device and guarantee a smooth and efficient operation. Observe the safety notes. Intended use Unit for cleaning the spray channels and lubricating the moving parts of dental transmission instruments, turbines, air motors and air-driven dental scalers.
  • Page 8 Introduction Production according to EU Directive The medical device meets the requirements of Regulation (EU) 2017/745 Responsibility of the manufacturer The manufacturer can only accept responsibility for the safety, reliability and performance of the medical device when it is used in compliance with the following directions: >...
  • Page 9: Unpacking

    2. Unpacking Have the Instructions for Use at Take out the service oil, cleaning hand. Take out the white and liquid, suction filter and the green funnels. optional accessories from the side compartments. Remove the basic adaptor. Lift out the Assistina in the fibre- board box.
  • Page 10: Scope Of Delivery

    3. Scope of delivery > W&H Assistina maintenance unit (REF 00030125, 00030140) > Suction filter, complete > Basic adaptor REF 02685000 > Funnel, white, REF 02689300 > Funnel, green, REF 02689400 > Air connection hose 2 m, REF 02697000, complete >...
  • Page 11: Safety Notes

    4. Safety notes Please observe the following instructions under all circumstances > Only set up the product in suffiently well ventilated rooms! > Keep the device closed during the maintenance process! GHS02 WARNING! H226 Flammable liquid and vapour > P210 Keep away from heat, hot surfaces, sparks, open flames and other ignition sources. No smoking. >...
  • Page 12: Description

    5. Description Front side Filler hole / Filler hole / storage container storage container for service oil for cleaning liquid Recess for Button to cleaning / adaptor maintenance process Fill level of Fill level of cleaning service oil liquid Indicator window Cover for service oil Waste disposal...
  • Page 13: Rear Side

    Description Rear side Suction filter Quick release coupling for compressed air...
  • Page 14: Start-Up

    6. Start-up > Service oil must be at room temperature when it is poured in. > Never reverse funnels (white to white, green to green) and store them in a dust-free environment! > Check the filling level indicator! Unscrew the two caps and remove the two transport seals (malfunction!)!
  • Page 15 Start-up Unscrew the cap from the Add 250 ml of cleaning liquid. white storage container. Insert the white funnel. Add 250 ml of service oil. Close the storage containers for the service oil and cleaning liquid. Unscrew the cap from the green storage container.
  • Page 16 Start-up General Push the quick release coupling of Fit the suction filter so that it 02697000 the connecting hose into the is leak-tight. coupling socket on the Assistina. Connect the compressed air line. The built-in air filter in the connecting hose is only a fine filter. It cannot clean unfiltered and possibly contaminated compressor air.
  • Page 17: Initial Filling

    Start-up Initial filling Pull the drip guard off the Press the start button about maintenance attachment and 15 times for 2 seconds with a 15 x store it safely. pause of 2 seconds between 2 secs. Close the cover. each actuation. If no fluid can be seen or detected on the waste disposal tray, the supply lines are not yet completely full.
  • Page 18 Start-up Place a fleece inlay in the Press the start button once for waste disposal tray. 2 seconds. 2 secs. Close the cover and waste disposal tray. Check that the fleece inlay is damp to ensure correct function. If it is not damp, repeat the process 2 to 3 times. Dispose of saturated fleece inlays as with cloths for surface disinfection.
  • Page 19 Start-up Function display Continuously monitor the function display! > Press the start button for 2 seconds. A green ball for the cleaning liquid appears in the right-hand window. > Release the start button. A green ball for the service oil appears in the left-hand window. After a few seconds, both balls will drop down again.
  • Page 20: Operation

    7. Operation Maintenance process > Wear protective clothing, saftey glasses, face mask and gloves. > Refer to the »hygiene and maintenance« instructions in the Instructions for Use of the transmission instruments. Transmission instruments with ISO coupling The maintenance program will be completed in about 35 seconds. >...
  • Page 21: Attaching / Removing The Adaptor

    Operation Attaching / removing the adaptor Turbines The maintenance program will be completed in about 35 seconds. > Use the basic adaptor REF 02685000 with standard 4-hole connection. > Pull down cover. W&H turbines: Turbines with 02685000 > 2-hole or 3-hole Borden Screw together the basic 04713200 adaptor and the adaptor.
  • Page 22: Maintenance Process

    Operation Maintenance process Push the appropriate adaptor Press the start button once onto the maintenance for about 2 seconds. Open 2 sec. attachment up to the the cover at the end of the limit stop. maintenance cycle. With its spray exit nozzle Press the green release directed downwards, push button of the maintenance...
  • Page 23 Operation Maintenance process Air motors and air-driven dental scalers The maintenance program will be completed in about 35 seconds. > Use the basic adaptor REF 02685000 with standard 4-hole connection. > Set the rotary ring of the air motor to clockwise or anticlockwise running (no maintenance will take place in the »Zero« setting). >...
  • Page 24: Lubrication Of Chucking System

    Operation Lubrication of chucking system FG contra-angle handpieces > Remove the rotary instrument. > Use the adaptor 02693000 > Wear protective clothing, saftey glasses, face mask and gloves. With its spray exit nozzle Press the start button once for directed downwards, push the about 2 seconds.
  • Page 25: Cleaning

    8. Cleaning Follow your local and national laws, directives, standards and guide lines for cleaning and disinfection. > Wear protective clothing, saftey glasses, face mask and gloves. Cleaning agents > Use only detergents with are intended for cleaning medical devices made of metal and plastic. W&H recommends external cleaning of the Assistina >...
  • Page 26: W&H Accessories

    9. W&H accessories Use only original W&H accessories / spare parts or accessories approved by W&H Optional accessories 02676400 Drip guard for maintenance attachment 02679000 Adaptor for removable contra-angle handpiece heads 02680600 Seal 02685000 4-hole basic adaptor 02689300 Funnel, white 02689400 Funnel, green 02690400...
  • Page 27 W&H accessories 02675200 Air filter for air connection hose 02697000 02695700 O-ring for maintenance attachment 02670500 Filter pack 02675910 500 ml W&H service oil 02680200 1000 ml W&H cleaning liquid 02680500 Pack of W&H service oil and W&H cleaning liquid 02690300 Initial provision of 5 layers of fleece 02694500...
  • Page 28: Maintenance

    10. Maintenance Regular testing of Assistina Regular testing of function and safety is necessary: With each application: > Check the function display Once a week: > Check the filling level displays. Once a year: > Carry out a function check (page 19). >...
  • Page 29: O-Rings Replacement

    Maintenance O-rings replacement > Replace damaged or leaking O-rings immediately. > Do not use sharp tools! Check the O-rings on the Firmly squeeze the O-ring maintenance attachment. between thumb and forefinger Change the O-rings in the event and pull it off. of visible defects or every two Push on a new O-ring.
  • Page 30: Filter Replacement

    Maintenance Filter replacement Renewing the filter pack > Do not touch the vents of the suction filter (risk of damage)! > Use the W&H Assistina for maintenance only when the suction filter with a fully serviceable filter pack is fitted. >...
  • Page 31: Air Filter Replacement

    Maintenance Air filter replacement Switch off the central air supply before replacing the air filter. > Replace the air filter whenever it is dirty or at least 1x per year. Press the ring downwards. At the Replace the air filter and same time, pull out the air re-connect the air hose (a, b).
  • Page 32: Servicing

    11. Servicing Returns Returns > If you have any questions, please contact your local dental dealers or one of our service points. (see page 37) Always return equipment in the original packaging. 2 transport seals No. 02680600 are attached here.
  • Page 33: Preparation Of Returns

    Servicing Preparation of returns Pull out the suction Fit a seal in each filler hole. filter horizontally. Screw on the two caps. Before emptying the first container, fit the transport seal on the second container. Disconnect the Fit the drip guard No. air connection hose.
  • Page 34: Technical Data

    12. Technical data Height: 223 mm Width: 190 mm Depth (with filter): 415 mm Weight: 2.70 kg Air consumption: approx. 60 Nl/min. Noise: 63 dB (A) (at 50 cm distance) Flow pressure: (pressure regulation provided integral automatic pressure regulator) Capacity: 250 ml W&H service oil F1, sufficient for approx.
  • Page 35: Disposal

    13. Disposal Ensure that the parts are not contaminated on disposal. Follow your local and national laws, directives, standards and guidelines for disposal. > Medical device > Packaging 35 35...
  • Page 36: Explanation Of Warranty Terms

    E xp la na ti on o f wa r ranty ter ms This W&H medical device has been manufactured with great care by highly qualified specialists. A wide variety of tests and controls guarantee faultless operation. Please note that claims under warranty can only be validated when all the directions in the Instructions for Use have been followed.
  • Page 37: Authorised W&H Service Partners

    Authorised W&H service partners Find your nearest W&H service partner at http://wh.com Simply go to the menu option »Service« for full details. Or simply scan the QR code.
  • Page 38 Manufacturer W&H Dentalwerk Bürmoos GmbH Ignaz-Glaser-Straße 53, 5111 Bürmoos, Austria Form No. 50458 AEN Rev. 019 / 09.03.2020 t +43 6274 6236-0, f +43 6274 6236-55 Subject to alterations office@wh.com wh.com...