Waltron 9065 Quick Reference Manual page 9

Luminescent dissolved oxygen analyzer calibration kit
Hide thumbs Also See for 9065:

Advertisement

Step
3.4
3.5
104-001-A.7 Released 06/14/2019
Description
Using the Arrow buttons select
O2 AUTO. Press ENTER button.
In the next screen select O2 LOW.
Press ENTER button.
Select 000000 ppb. Press ENTER
button.
Calibration will begin. Wait
approximately 2 minutes. Do not
disturb the analyzer or the
calibration equipment.
Key Points