Introduction And Setup - Waltron 9065 Quick Reference Manual

Luminescent dissolved oxygen analyzer calibration kit
Hide thumbs Also See for 9065:

Advertisement

Step
1

INTRODUCTION AND SETUP

1.1
1.2
1.3
1.4
104-001-A.7 Released 06/14/2019
Calibration is Due!
Description
Read this manual. Ensure you
have an understanding of the key
points listed in the right hand
column.
Make sure you have all items
necessary to complete calibration.
Inventory your calibration kit
(Calibration Kit p/n K6000-101).
Determine if the Dissolved Oxygen
Analyzer requires calibration. If
the screen is flashing and a
triangle appears in the upper right
corner, your 9065 sensor needs
calibration.
Turn off sample flow to the
analyzer.
Key Points
A calibration beaker is provided
with each 9065 analyzer order.
Contact Waltron for additional
beakers p/n K5000-039.
Each sensor comes with an
internal light pulse counter.
When the light pulse counter
reaches 1 million your analyzer
requires calibration.